A comparison between visual, robotic, and intracoronary imaging lesion length assessment during robotic-assisted percutaneous coronary intervention.

James Leung, John French, James Xu, Kathryn Wales, Hashim Kachwalla, Krishna Kaddapu, Sarah Fares, Viet Dang,

Department of Cardiology, Liverpool Hospital, Sydney, New South Wales, Australia; South West Sydney Clinical School, University of New South Wales, Australia. Email: james.t.leung@unsw.edu.au. · Department of Cardiology, Liverpool Hospital, Sydney, New South Wales, Australia; South West Sydney Clinical School, University of New South Wales, Australia.

Key points

From the paper's conclusion
  • Visual assessment of lesion length often underestimates true lesion length. R-PCI may provide more accurate and reproducible lesion measurements, with better correlation and agreement with ICI measurements than visual estimation.

Abstract

OBJECTIVES: Inaccurate lesion length assessment can lead to adverse clinical outcomes after percutaneous coronary intervention (PCI). Robotic-assisted PCI (R-PCI) may provide more accurate lesion assessment. The authors aimed to compare the accuracy of visual, robotic, and intracoronary imaging-derived lesion length assessment. METHODS: Sixty-three patients (69 lesions) who underwent R-PCI using the CorPath GRX (Siemens Healthineers) system were analyzed. The operator provided a visual estimate of lesion length, which was compared to the measurement using the robotic system. In cases in which intracoronary imaging (ICI) was performed (n=21), the visual and robotic estimates were compared against the ICI measurements, which were considered the gold standard. RESULTS: The median visual length measurement was 23.0 mm (IQR: 18.0-28.0) and the median robotic visual length measurement was 24.0 mm (IQR: 18.7-33.0 mm). In only 10.1% of lesions did the visual and robotic estimates agree. In 66.7% of the lesions, the visual estimate was shorter than the robotic measurement, whereas the visual estimate was longer than the robotic measurement in 23.2% of lesions. Using a threshold of greater than 5 mm too long/short, 22 lesions (32%) were inaccurately measured. Both visual (r=0.82) and robotic (r=0.97) estimates had good correlation with ICI measurements; however, this difference in correlation was statistically significant (P = .004). Visual assessment consistently underestimated lesion length compared with ICI (mean difference -4.2 ± 6.9 mm). CONCLUSIONS: Visual assessment of lesion length often underestimates true lesion length. R-PCI may provide more accurate and reproducible lesion measurements, with better correlation and agreement with ICI measurements than visual estimation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers