Displayed CTDIvolume accuracy in ultra-low-dose CT: Limitations of current IEC criteria and a proposal for new constancy benchmarks.

T Shohji, R Tomoika, T Kobayashi

Department of Radiology, The Jikei University Katsushika Medical Center, 6-41-2 Aoto, Katsushika-ku, Tokyo, 125-8506, Japan. Electronic address: t_shouji@jikei.ac.jp. · Department of Radiology, The Jikei University Katsushika Medical Center, 6-41-2 Aoto, Katsushika-ku, Tokyo, 125-8506, Japan. Electronic address: r_tomioka@jikei.ac.jp.

Computational

Key points

From the paper's conclusion
  • The ±1 mGy absolute constancy criterion loses effectiveness in the ultra-low-dose regime, creating a critical latent risk for undetected equipment degradation. International QC standards urgently require revision to ensure reliable dose monitoring.

Abstract

INTRODUCTION: Rapid adoption of ultra-low-dose CT protocols (∼1-2 mGy) used in paediatric cardiac CT examinations challenges current quality control (QC). The IEC 61223-3-5 constancy criterion for CTDIvol display accuracy ("±20% or ±1 mGy") permits up to ±100% relative error at 1.0 mGy, undermining patient dose management and early detection of equipment anomalies. This study evaluates ultra-low-dose display accuracy and proposes revised QC benchmarks. METHODS: CTDIvol accuracy was evaluated across two state-of-the-art CT systems using a cylindrical polymethyl methacrylate (PMMA) phantom with a 32 cm-diameter and a 15 cm-length and a calibrated semiconductor detector. Measurements ranged from 10 to 0.5 mGy (120, 100, and 80 kV), utilizing synchronized and asynchronous rotational exposure methods to ensure geometric reproducibility. RESULTS: Measured doses remained within a ±20% relative tolerance across a tested range from 0.5 mGy to 10.3 mGy, proving precise control is achievable in the ultra-low-dose range. However, system-specific discrepancies were identified, with manufacturers showing distinct overestimation or underestimation tendencies. Crucially, applying the ±1 mGy absolute tolerance below 5.0 mGy permitted exponentially increasing relative errors, effectively nullifying fundamental QC objectives. CONCLUSION: The ±1 mGy absolute constancy criterion loses effectiveness in the ultra-low-dose regime, creating a critical latent risk for undetected equipment degradation. International QC standards urgently require revision to ensure reliable dose monitoring. IMPLICATIONS FOR PRACTICE: We propose establishing a "low-dose boundary value" at 5.0 mGy. Above this threshold, the existing ±1 mGy criterion applies; below it, a uniform ±20% relative tolerance is strictly enforced. This provides a robust, backward-compatible QC framework for low-dose CT, highlighting the need to recognize manufacturer-dependent display uncertainties during protocol optimization.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Displayed CTDIvolume accuracy in ultra-low-dose CT: Limitations of current IEC criteria and a proposal for new constancy benchmarks — Science Skim