Displayed CTDIvolume accuracy in ultra-low-dose CT: Limitations of current IEC criteria and a proposal for new constancy benchmarks.
T Shohji, R Tomoika, T Kobayashi
Department of Radiology, The Jikei University Katsushika Medical Center, 6-41-2 Aoto, Katsushika-ku, Tokyo, 125-8506, Japan. Electronic address: t_shouji@jikei.ac.jp. · Department of Radiology, The Jikei University Katsushika Medical Center, 6-41-2 Aoto, Katsushika-ku, Tokyo, 125-8506, Japan. Electronic address: r_tomioka@jikei.ac.jp.
Key points
From the paper's conclusion- The ±1 mGy absolute constancy criterion loses effectiveness in the ultra-low-dose regime, creating a critical latent risk for undetected equipment degradation. International QC standards urgently require revision to ensure reliable dose monitoring.
Abstract
INTRODUCTION: Rapid adoption of ultra-low-dose CT protocols (∼1-2 mGy) used in paediatric cardiac CT examinations challenges current quality control (QC). The IEC 61223-3-5 constancy criterion for CTDIvol display accuracy ("±20% or ±1 mGy") permits up to ±100% relative error at 1.0 mGy, undermining patient dose management and early detection of equipment anomalies. This study evaluates ultra-low-dose display accuracy and proposes revised QC benchmarks. METHODS: CTDIvol accuracy was evaluated across two state-of-the-art CT systems using a cylindrical polymethyl methacrylate (PMMA) phantom with a 32 cm-diameter and a 15 cm-length and a calibrated semiconductor detector. Measurements ranged from 10 to 0.5 mGy (120, 100, and 80 kV), utilizing synchronized and asynchronous rotational exposure methods to ensure geometric reproducibility. RESULTS: Measured doses remained within a ±20% relative tolerance across a tested range from 0.5 mGy to 10.3 mGy, proving precise control is achievable in the ultra-low-dose range. However, system-specific discrepancies were identified, with manufacturers showing distinct overestimation or underestimation tendencies. Crucially, applying the ±1 mGy absolute tolerance below 5.0 mGy permitted exponentially increasing relative errors, effectively nullifying fundamental QC objectives. CONCLUSION: The ±1 mGy absolute constancy criterion loses effectiveness in the ultra-low-dose regime, creating a critical latent risk for undetected equipment degradation. International QC standards urgently require revision to ensure reliable dose monitoring. IMPLICATIONS FOR PRACTICE: We propose establishing a "low-dose boundary value" at 5.0 mGy. Above this threshold, the existing ±1 mGy criterion applies; below it, a uniform ±20% relative tolerance is strictly enforced. This provides a robust, backward-compatible QC framework for low-dose CT, highlighting the need to recognize manufacturer-dependent display uncertainties during protocol optimization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dose-aware 5-star image quality rating in CT: Findings from the IAEA-MGH study.European journal of radiology · 2026
- Ultra-Low-Dose Noise-Free Technology to Reduce Radiation Exposure During Coronary AngiographyCirculation Cardiovascular Interventions · 2026
- Assessment of image quality and radiation dose in low-dose, low-contrast computed tomography coronary angiogram: a systematic review and meta-analysis.Clinical radiology · 2026
- Anatomy-Weighted CTDIvol from Routine CT Metadata: A Patient-Specific, Multi-Vendor Study Using Deep-Learning Segmentation.Tomography (Ann Arbor, Mich.) · 2026
- Dose optimization in contrast-enhanced CT brain imaging using low tube voltage and reduced iodine concentration: a comparative study.BMC medical imaging · 2026
- Evaluation of iodine detection and quantification limits in photon-counting CT-a phantom study.European radiology experimental · 2026