Comparison of dosimetric verification results and plan complexity for treatment plans generated in Ethos 1.0 and Ethos 2.0 systems.
Paulina Wesolowska, Joanna Iwanowska-Hanke, Anna Paciorkiewicz, Sandra Maluszczak, Mikołaj Tarchalski, Bartłomiej Sadowski, Marta Fillmann, Agnieszka Walewska
Medical Physics Department, Maria Skłodowska-Curie National Research Institute of Oncology, Warsaw, Poland. paulina.wesolowska@nio.gov.pl. · Medical Physics Department, Maria Skłodowska-Curie National Research Institute of Oncology, Warsaw, Poland.
Key points
From the paper's conclusion- The upgrade preserved quality-assurance compliance and produced simpler, lower--monitor-unit plans, but the systematic -1.5% offset between calculated and measured point dose persisted, showing the new MLC model alone does not remove it.
Abstract
INTRODUCTION: The Varian Ethos adaptive radiotherapy system was upgraded from version 1.0 to 2.0, introducing an enhanced multi-leaf collimator (MLC) dose calculation model and a new optimization engine. We evaluated whether this affected agreement between calculated and measured dose, and plan complexity. MATERIAL AND METHODS: Twenty-four plans (14 intensity-modulated radiotherapy [IMRT], 10 volumetric modulated arc therapy [VMAT]; 26 measurement configurations) covering 10 anatomical sites were analyzed in three matched configurations: original Ethos 1.0 plans (E1.0), the same plans recalculated in Ethos 2.0 with the new MLC model only (E1.0→2.0), and plans re-optimized in Ethos 2.0 (E2.0). Quality assurance comprised ionization chamber point-dose measurement and global gamma analysis on an ArcCHECK array at 3%/3 mm, 3%/2 mm and 2%/2 mm (10% threshold). Complexity was quantified with UCoMX v1.1; groups were compared with the Wilcoxon signed-rank test. RESULTS: Mean signed chamber deviation was -1.59 ± 0.54% (E1.0), -1.54 ± 0.57% (E1.0→2.0) and -1.39 ± 0.63% (E2.0), with no significant differences (p > 0.6); all met the ±3% criterion. Gamma pass rates at 3%/3 mm were 99.2 ± 0.9%, 98.7 ± 1.0% (p < 0.001) and 98.6 ± 2.1% (p = 0.059). Re-optimized plans used 12% fewer monitor units (1,169 ± 395 vs. 1,338 ± 471, p < 0.001) and were simpler (MCSv 0.31 to 0.35; SAS10 mm 0.27 to 0.20; both p < 0.001). CONCLUSION: The upgrade preserved quality-assurance compliance and produced simpler, lower--monitor-unit plans, but the systematic -1.5% offset between calculated and measured point dose persisted, showing the new MLC model alone does not remove it.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dosimetric Validation of Ethos Online Adaptive Radiotherapy for Prostate Cancer: An End-to-End Phantom Study.Life (Basel, Switzerland) · 2026
- Platform-specific comparison of IMRT and VMAT angular resolution for Ethos-based online adaptive radiotherapy in cervical cancer.Frontiers in oncology · 2026
- Assessing the feasibility of uTPS for volumetric modulated arc therapy treatment plan optimization of dual-layer multileaf collimator linear accelerators.Radiation oncology (London, England) · 2026
- Assessing the impact of knowledge-based planning on dosimetric performance in adaptive prostate stereotactic body radiation therapy.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Configuration of a Monte Carlo-Based Secondary Plan Verification Tool: An Analytical Approach to User-Specific Parameter Tuning.Biomedical physics & engineering express · 2026
- Exploring a hybrid workflow on current adaptive radiotherapy systems.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026