Risk Factors for Cement Leakage into Segmental Veins during Posterior Spinal Fusion Using Fenestrated Pedicle Screws.
Hayato Futakawa, Hiroto Makino, Shoji Seki, Katsuhiko Kamei, Yushi Yashima, Yasuhito Yahara, Yoshiharu Kawaguchi
Department of Orthopaedic Surgery, Toyama University, Toyama, Japan.
Key points
From the paper's conclusion- Proximity of the screw tip to the anterior and lateral walls, as well as excessive cement injection relative to vertebral body size, significantly increases the risk of leakage. To mitigate this risk, screws should be placed centrally within the vertebral body and at a safe distance from the cortical boundaries. The amount of injected cement should also be determined according to the volume of the individual vertebral body. These findings provide practical guidance for safer screw placement in spinal fusion using fenestrated screws.
Abstract
INTRODUCTION: Although fenestrated screws enhance pull-out strength, cement leakage into segmental veins carries a risk of pulmonary embolism. This study aimed to identify factors that contribute to cement leakage during posterior spinal fusion with fenestrated screws. METHODS: We conducted a retrospective analysis of 32 patients who underwent posterior spinal fusion with fenestrated pedicle screws. A total of 147 screws were evaluated using postoperative computed tomography (CT) scans. Risk factors were assessed by measuring vertebral CT values, the volume of injected cement and its ratio to vertebral body volume, and screw positioning parameters, including deviation and distance to the anterior and lateral walls. RESULTS: Cement leakage occurred in 75% of patients (24 of 32) and in 35% of screws (52 of 147). All leakages were classified as leakage into segmental veins. No symptomatic pulmonary embolism was observed. The thoracic spine exhibited a higher leakage rate than the lumbar spine (47.5% vs. 28.4%, p<0.05). Leakage was significantly associated with shorter distances from the screw tip to the anterior (5.1 mm vs. 7.8 mm) and lateral (4.8 mm vs. 7.3 mm) vertebral walls, as well as a higher cement-to-vertebral-volume ratio (7.1% vs. 4.7%, p<0.01). Leakage also correlated with higher CT values in the posterior vertebral region (126 Hounsfield Unit [HU] vs. 103 HU; p<0.05). Screw deviation showed no association. CONCLUSIONS: Proximity of the screw tip to the anterior and lateral walls, as well as excessive cement injection relative to vertebral body size, significantly increases the risk of leakage. To mitigate this risk, screws should be placed centrally within the vertebral body and at a safe distance from the cortical boundaries. The amount of injected cement should also be determined according to the volume of the individual vertebral body. These findings provide practical guidance for safer screw placement in spinal fusion using fenestrated screws.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient Selection and Reoperation After Cement-Augmented Pedicle Screw Fixation in the Thoracolumbar Spine: A Single-Center Cohort.Journal of clinical medicine · 2026
- Risk factors associated with bone cement leakage following percutaneous kyphoplasty: a retrospective analysisEuropean journal of medical research · 2025
- Risk Factors for Bone Cement Leakage and Re-fracture Following PKP in Patients With Thoracolumbar Spine Fractures.Annali italiani di chirurgia · 2026
- Endovascular retrieval of bone cement entrapped in an inferior vena cava filter after percutaneous vertebroplasty: a case report.Frontiers in surgery · 2026
- Procedural Complications and Implant Cutout of the SpineJack®-System in Thoracolumbar Vertebral Fractures: A Retrospective Analysis of 204 Implants from a Level-1 Trauma Center.Global spine journal · 2026
- Combined unilateral biportal endoscopy-assisted posterior reconstruction and transforaminal endoscopic cave-in decompression for ventral thoracic cement leakage: a case report and technical noteBMC Surgery · 2026