Technical Feasibility of a Multimodal Electro-Optical and Infrared Imaging Pipeline for Postoperative Wound Assessment After Total Knee Arthroplasty: Formative Evaluation Study.
Tej S Ranade, Emily Hu, Kadambari Beelwar, Sandeep Boddeda, Praharsha Mulpur, Adarsh Annapareddy, A V Gurava Reddy
Truss Health, Inc., Las Vegas, NV, United States. · Sunshine Bone and Joint Institute, Krishna Institute of Medical Sciences, Begumpet, Telangana, India.
Key points
From the paper's conclusion- In this formative study, a multimodal EO and infrared imaging pipeline demonstrated technical feasibility for postoperative wound assessment following TKA, providing noncontact discrimination of infection-related complications that compared favorably with established adjuncts. This work provides proof-of-concept for multimodal imaging-based assessment and defines the technical requirements for the prospective, multicenter external validation needed to establish clinical utility.
Abstract
BACKGROUND: Early detection and monitoring of wound-related complications after total knee arthroplasty (TKA) are critical to optimize outcomes and preserve implants. Conventional postoperative monitoring relies on in-person assessment and patient-reported symptoms, which may delay recognition. Noncontact imaging acquired in the clinic or at home offers a potential route to earlier and more consistent assessment. OBJECTIVE: This study aimed to evaluate the technical feasibility and performance of a multimodal AI system integrating electro-optical (EO) and infrared imaging for postoperative wound assessment following TKA and to explore clinical factors associated with early complications. METHODS: We conducted a single-center prospective cohort study of 749 patients undergoing primary TKA. Patients underwent standardized multimodal data collection at scheduled postoperative visits. EO-based models localized surgical landmarks and segmented complication-oriented tissue states; a fine-tuned vision-language model provided protocol verification and contextual visual reasoning. EO landmarks mapped to thermal space via deterministic alignment, followed by infrared region-of-interest segmentation and thermal classification. Each module was evaluated on a defined held-out set. The thermal classifier was assessed out-of-fold under 5-fold cross-validation with folds grouped by patient, in knees imaged on postoperative day 14 or later. Discrimination for clinician-adjudicated infection-related complications served as the primary outcome. Comorbidities and baseline laboratory values were explored for associations with complications. RESULTS: Among 749 patients, 26 (3.5%) had a clinician-adjudicated infection-related complication. On held-out evaluation, the EO landmark detector reached a sensitivity of 0.967 (95% CI 0.954-0.979) and positive predictive value of 0.955 (95% CI 0.940-0.970), and the wound-tissue segmentation model reached a sensitivity of 0.845 (95% CI 0.833-0.857) and positive predictive value of 0.759 (95% CI 0.744-0.772). Infrared region-of-interest extraction isolated the surgical field with a mean intersection-over-union of 0.985 (SD 0.008; 95% CI 0.984-0.986) across 178 held-out patients. The trained thermal classifier discriminated infected scans with an area under the receiver operating characteristic curve (AUROC) of 0.978 (95% CI 0.959-0.992) in out-of-fold, patient-grouped cross-validation. At a representative operating point, sensitivity was 0.808 (95% CI 0.65-0.92), specificity 0.974 (95% CI 0.96-0.99), positive predictive value 0.618 (95% CI 0.49-0.78), and negative predictive value 0.990 (95% CI 0.98-1.00). Exploratory analyses showed nominal associations of chronic kidney disease and diabetes mellitus with postoperative complications that did not remain significant after multiplicity correction. CONCLUSIONS: In this formative study, a multimodal EO and infrared imaging pipeline demonstrated technical feasibility for postoperative wound assessment following TKA, providing noncontact discrimination of infection-related complications that compared favorably with established adjuncts. This work provides proof-of-concept for multimodal imaging-based assessment and defines the technical requirements for the prospective, multicenter external validation needed to establish clinical utility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dynamic infrared thermography and computed tomography angiography fusion for DIEP flap perforator selection: a novel measurement systemarXiv · 2026
- Manual vs. Automatic Segmentation in Infrared Thermography: Inter-Rater Reliability and a Longitudinal Single-Case Feasibility Illustration of Agreement for Musculoskeletal Rehabilitation Monitoring.Journal of functional morphology and kinesiology · 2026
- Multimodal Deep Learning for Wound Healing Prediction.European journal of pharmaceutical sciences : official journal of the European Federation for Pharmaceutical Sciences · 2026
- Detecting Incisional Surgical Site Infections on Wound Images Through Deep Learning.JAMA surgery · 2026
- Imaging Informatics for Thermal Ablation: CT Thermometry-Based Tissue Injury Mapping.IEEE journal of biomedical and health informatics · 2026
- Automated Infrared Thermography and Machine Learning for Detecting Physiological Thermal Patterns in Lumbar Radiculopathy: An Exploratory Pilot Study.Biomimetics (Basel, Switzerland) · 2026