Magnetic Resonance Imaging-Based Measurement of Pediatric Head Circumference by a Clinician-in-the-Loop 3D Surface Tool: Development and Validation Study.
Leonie Dittrich, Jan Philipp Adam, Ute von Jan, Eva Bültmann, Dominik Wolff
Peter L. Reichertz Institute for Medical Informatics of TU Braunschweig and Hannover Medical School, Hannover Medical School, Carl-Neuberg-Straße 1, Hannover, Lower Saxony, 30625, Germany, 49 5115325372. · Institute of Diagnostic and Interventional Neuroradiology, Hannover Medical School, Hannover, Lower Saxony, Germany.
Key points
From the paper's conclusion- The clinician-guided workflow enabled MRI-based HC measurement in all examined cases and showed small interrater differences. Delaunay-based interpolation was associated with closer agreement with tape-based measurements in the small subgroup with cranial surface irregularities, while both contour variants performed similarly in examinations without irregularities. These findings are preliminary because of the small cohort, the single tape-based reference measurement per patient, and the 5-case irregularity subgroup. Larger studies using repeated reference measurements and broader cranial morphologies are needed to validate the workflow.
Abstract
BACKGROUND: Head circumference (HC) is a routine clinical parameter in pediatric patients indicating brain growth and neurodevelopment. Although usually obtained with a tape measure, this method is operator-dependent and prone to variability. Magnetic resonance imaging (MRI) provides a retrospective alternative, but existing imaging-based approaches can be limited by predefined measurement planes, restricted clinician review, and cranial surface artifacts. OBJECTIVE: This study aimed to develop and evaluate a clinician-guided software tool for measuring pediatric HC from routinely acquired 3D MRI examinations. METHODS: We developed a desktop application in Python using PySide6, the Insight Segmentation and Registration Toolkit, and the Visualization Toolkit for MRI-based HC measurement. The workflow included outer skin surface reconstruction, placement of 3 landmarks to define a freely orientable measurement plane, contour extraction, and circumference calculation. Optional Delaunay-based interpolation was implemented to reduce the influence of local cranial surface irregularities. In this retrospective evaluation, the tool was applied to 23 pediatric MRI examinations, including 18 without and 5 with clinically identified cranial surface irregularities. Two clinicians independently performed 3 repeated software measurements per examination. Median software measurements were compared with a single tape-based HC measurement obtained on the date of MRI. Agreement was assessed using Bland-Altman analysis, mean absolute error (MAE), proportional bias analysis, and a linear mixed-effects model. Interrater differences and preliminary usability feedback were also evaluated. RESULTS: Software-based HC measurement was feasible in all 23 examinations. In cases without cranial surface irregularities, the raw and interpolated contours showed similar agreement with tape-based measurements (bias 0.415 cm and 0.410 cm; MAE 0.429 cm and 0.423 cm, respectively). In the 5 examinations with irregularities, the intersection contour showed a greater bias of 1.420 cm and wider limits of agreement (LoA; -0.182 to 3.021 cm), whereas the interpolated contour showed a lower bias of 0.448 cm and narrower LoA (-0.119 to 1.014 cm). The estimated coefficient for the contour-by-subgroup interaction was -0.967 cm (SE 0.195 cm, 95% CI -1.349 to -0.584 cm; P<.001). No statistically significant proportional trend was detected, although the analysis had limited power to exclude proportional bias. Interrater differences were small. The 2 clinicians reported System Usability Scale scores of 92.5 and 95.0, representing preliminary descriptive feedback. CONCLUSIONS: The clinician-guided workflow enabled MRI-based HC measurement in all examined cases and showed small interrater differences. Delaunay-based interpolation was associated with closer agreement with tape-based measurements in the small subgroup with cranial surface irregularities, while both contour variants performed similarly in examinations without irregularities. These findings are preliminary because of the small cohort, the single tape-based reference measurement per patient, and the 5-case irregularity subgroup. Larger studies using repeated reference measurements and broader cranial morphologies are needed to validate the workflow.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Charting neuroimaging-based head circumference and cranial shape across development.bioRxiv : the preprint server for biology · 2026
- ClinSeg: Robust Brain Segmentation for Clinically Acquired Pediatric MRI.medRxiv : the preprint server for health sciences · 2026
- Automated Cerebrospinal Fluid Flow Quantification in Hydrocephalus Using Phase‐Contrast Magnetic Resonance ImagingNMR in Biomedicine · 2026
- Improved SNR and Image Quality of Pediatric MRI with a Shape Conforming High-Density Posterior Cushion Coil.Bioengineering (Basel, Switzerland) · 2026
- A Triple-Measure Model for Estimating Cranial-Brain Size Across Age GroupsbioRxiv · 2026
- Why probe registration is crucial for infant fNIRS and HD-DOT? An evaluation across 49 newbornsbioRxiv · 2026