Correcting Differences Between Dexcom G6 Pro and FreeStyle Libre Pro Glucose Readings: Retrospective Paired-Device Study.
Faris Raza, Peter Yiğitcan Washington
Artificial Intelligence and Computational Drug Discovery and Development (AICD3), Department of Bioengineering and Therapeutic Sciences, University of California, San Francisco, 1700 Fourth St, San Francisco, US. · Division of Clinical Informatics and Digital Transformation (DoC-IT), Department of Medicine, University of California, San Francisco, San Francisco, US.
Key points
From the paper's conclusion- In exploratory analyses, these sensors were not interchangeable, and no correction was best for every endpoint. Because Dexcom averaged above and Libre below blood glucose in matched samples, correcting one toward the other improves comparability, not necessarily accuracy. Studies combining CGM data should name the target sensor and match the method to the endpoint: glucose values, glycemic zones, or glycemic stage.
Abstract
BACKGROUND: Continuous glucose monitors (CGMs) from different manufacturers can read differently for the same person. Studies combining data across sensors can correct readings toward one sensor (the target) for comparability. However, it remains unclear how much reducing glucose error improves agreement on clinical endpoints such as time in range (TIR) and glucose zones. Sensor differences may also affect the consistency and accuracy of glycemic stage predictions. OBJECTIVE: To (1) compare Dexcom and FreeStyle Libre with each other and with capillary and laboratory glucose, (2) evaluate how well corrected readings match the target's glucose, TIR, and zones, and (3) assess whether richer CGM features improve cross-sensor diabetes stage classification without device identity. METHODS: In CGMacros, 45 adults, staged by glycated hemoglobin (15 healthy, 16 prediabetes, 14 type 2 diabetes [T2D]), simultaneously wore Dexcom G6 Pro and FreeStyle Libre Pro. Corrections used scaling, ridge regression, or a temporal convolutional network, fitted on other participants (population) or each person's first days (personalized) and tested on later days. We compared elastic net stage classifiers using 22 CGM features or mean glucose, with or without device identity, on both sensors in unseen participants. RESULTS: Libre averaged 32.37 mg/dL lower than Dexcom (95% CI 27.50-37.07). In matched samples, Dexcom averaged above and Libre below finger stick glucose (+16.51 and -14.33 mg/dL). External matched samples agreed in direction (Dexcom G6 in AI-READI, FreeStyle Libre H in ShanghaiT2DM). Libre showed 19.94 percentage points less TIR than Dexcom in healthy participants but 14.08 more in T2D. On Libre, TIR was higher in T2D than in healthy participants (82.33% vs 76.51%). Post hoc, one uniform shift of Dexcom readings could explain most of this opposite pattern. Toward Dexcom, personalized temporal ridge gave the lowest mean absolute error (10.73 mg/dL) and cut daily TIR error from 15.13 to 3.46 points, whereas personalized scaling detected zones better (balanced accuracy 77.9% vs 67.3%), at lower positive predictive value. Without device identity, a mean glucose model staged 30 of 45 participants differently across sensors; device identity reduced this to 13, and 22 CGM features without it to 12. Agreement did not imply correct staging. Neither model clearly discriminated better. Post hoc, adding these features to age, body mass index, and gender increased held-out discrimination in this exploratory internal evaluation (area under the curve gain 0.130; 95% CI 0.019-0.239, conditional on saved predictions). CONCLUSIONS: In exploratory analyses, these sensors were not interchangeable, and no correction was best for every endpoint. Because Dexcom averaged above and Libre below blood glucose in matched samples, correcting one toward the other improves comparability, not necessarily accuracy. Studies combining CGM data should name the target sensor and match the method to the endpoint: glucose values, glycemic zones, or glycemic stage.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Differences in the Glucose Management Indicator Between Continuous Glucose Monitoring Systems.Diabetes, obesity & metabolism · 2026
- Continuous glucose monitoring using the FreeStyle Libre Pro iQ in presymptomatic type 1 diabetes.Diabetologia · 2026
- Clinical accuracy of Dexcom G6 and G7 continuous glucose monitors in hospitalized pediatric patients with type 1 diabetes: a real-world studyAnnals of Pediatric Endocrinology & Metabolism · 2026
- A Comparative Analysis Between Original and Updated Glucose Management Indicator and Glycated Haemoglobin Across Continuous Glucose Monitoring Sensors in Type 1 Diabetes.Diabetes, obesity & metabolism · 2026
- Comparison of the Continuous Glucose Monitoring Profiles of Four Glucose-Lowering Medications in the GRADE Randomized TrialDiabetes Care · 2026
- Use of Continuous Glucose Monitoring Indicators to Guide Dose Titration of Weekly Basal Insulin Efsitora Alfa in Insulin-Naive and Basal Insulin-Experienced Adults with Type 2 Diabetes: A Retrospective Analyis.Diabetes technology & therapeutics · 2026