Impact of restoration design and material on the mechanical behavior of implant-supported monolithic CAD-CAM single crowns: An in vitro study.
Shaza Bishti, Rodi Bingöl, Stefan Wolfart, Taskin Tuna
Department of Prosthodontics RAK College of Dental Sciences, RAK Medical & Health Sciences University, Ras AlKhaimah, United Arab Emirates. · Department of Prosthodontics and Biomaterials, Centre for Implantology, Uniklinik RWTH Aachen, Germany.
Key points
From the paper's conclusion- Within study limitations, ACD did not increase screw loosening compared with CCD after cyclic loading. All designs and materials tolerated loads exceeding typical physiological masticatory forces.
Abstract
PURPOSE: To evaluate the effects of crown design and restorative material on screw loosening and load-bearing performance of monolithic single-implant restorations after cyclic loading. METHODS: Sixty implant-supported molar crowns were assigned to a conventional crown design (CCD; n=30), with a centrally positioned implant, or an alternative crown design (ACD; n=30), with a distally positioned implant and a premolar crown with cantilever. Each design was divided by restorative material (n=10): monochromatic zirconia (Zr_3), multilayer zirconia (Zr_3/5), and lithium disilicate (LS2). Crowns were fixed to titanium bases (35 Ncm) and underwent thermomechanical loading (98 N, 1,200,000 cycles). Occlusal load was applied centrally on CCD crowns and eccentrically on the cantilever in ACD groups. Reverse torque values before and after loading were recorded in Ncm, and reverse torque differences were calculated. Specimens were retightened (35 Ncm) and loaded to failure. Data were analyzed using two-way analysis of variance. RESULTS: No failures occurred during chewing simulation. ACD showed significantly lower reverse torque difference than CCD (6.72 ± 1.62 vs. 8.13 ± 1.70 Ncm; P = 0.002). CCD fracture loads were 6693 ± 738 N (Zr_3), 6969 ± 659 N (Zr_3/5), and 2330 ± 330 N (LS2). ACD-Zr_3/5 and ACD-Zr_3 showed abutment bending at 914.29 ± 212.92 N and 835.97 ± 70.91 N, respectively, while ACD-LS2 showed abutment bending followed by crown fracture at 1255.60 ± 195.88 N. CONCLUSIONS: Within study limitations, ACD did not increase screw loosening compared with CCD after cyclic loading. All designs and materials tolerated loads exceeding typical physiological masticatory forces.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Angulated Screw Channel (ASC) design on the fracture resistance of implant-supported monolithic zirconia and lithium disilicate crowns: an in vitro study.BMC oral health · 2026
- Fracture Load in Monolithic Multilayer Zirconia Crowns From Three Manufacturers: an in Vitro Study.Journal of dental research, dental clinics, dental prospects · 2026
- A Comparative Finite Element Study on the Biomechanical Effect of Implant Crown Screw Access Channel Filled by Direct vs. Indirect Restoration MaterialsProsthesis · 2026
- Tensile Retention of Lithium Disilicate and Zirconia Crowns Cemented to One-Piece Zirconia Implants: A Pilot In Vitro Study of Cementation Protocol, Resin Cement, and Micro-CT Cement Morphology.Materials (Basel, Switzerland) · 2026
- Mechanical behavior of endodontically treated molars restored using two restorative approaches and two CAD/CAM materials: an in-vitro study.BMC oral health · 2026
- Thermo-mechanical loading of innovative "Mini-skirt" preparation design versus conventional full-coverage maxillary central incisor preparation design: an in vitro study.BMC oral health · 2026