Effects of Dental Metal Implant Number and Configuration on PET/CT Quantification and Image Quality: A Phantom-Based Study.
Emine Tuna Akdogan, Sinem Coşkun, Nami Yeyin, Rabia Lebriz Uslu Beşli, Mustafa Demir
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Istanbul Kent University, Istanbul Türkiye. · Department Nuclear Medicine, Cerrahpasa Medicine Faculty, Istanbul University - Cerrahpasa, Istanbul, Türkiye.
Key points
From the paper's conclusion- Dental implants may affect PET/CT quantification, particularly for lesions adjacent to multiple implants; SUV measurements in these regions should be interpreted with caution.
Abstract
INTRODUCTION/OBJECTIVES: Dental metal implants can generate artifacts in PET/CT imaging, particularly in the head and neck region, affecting quantitative accuracy. This study evaluated variations in standardized uptake value (SUV), contrast (C), and contrast-to-noise ratio (CNR) across predefined dental implant configurations differing in both implant number and spatial arrangement and assessed their effects on simulated tongue, nasopharyngeal, and laryngeal lesions. METHODS: A custom-designed head and neck phantom containing extracted human teeth and up to eight titanium dental implants was used. Simulated tongue, nasopharyngeal, and laryngeal tumors were filled with F-18 FDG. PET/CT scans were acquired using seven implant configurations (0-8 implants) and reconstructed as attenuation-corrected (AC), AC with metal artifact reduction (AC+MAR), and non-attenuation-corrected (NoAC) images. SUV, contrast, and CNR were calculated using standardized ROIs. RESULTS: Compared with the implant-free configuration, tongue tumor SUV decreased by 11% and 9.1% on AC and AC+MAR images, respectively, at eight implants. Nasopharyngeal and laryngeal SUV showed only minimal descriptive change with increasing implant number and configurations on AC+MAR images. Contrast and CNR followed similar trends on AC and AC+MAR images, whereas NoAC images consistently showed substantially lower contrast and CNR regardless of implant number. DISCUSSION: Although MAR reduced implant-induced artifacts and partially attenuated SUV underestimation, it did not eliminate the configuration-dependent trend observed across the predefined implant arrangements. Lesions farther from the implants appeared less affected after MAR; however, this should be regarded as a hypothesis-generating observation from a single phantom experiment, requiring validation in patient studies and phantom models with varying implant-to-lesion distances. CONCLUSION: Dental implants may affect PET/CT quantification, particularly for lesions adjacent to multiple implants; SUV measurements in these regions should be interpreted with caution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Photon-Counting Detector CT for Imaging Around Orthopedic and Spinal Metallic Implants: A Systematic Review.Cureus · 2026
- Evaluation of virtual monochromatic imaging combined with metal artifact reduction in low-dose dual-energy computed tomography: a phantom study.Quantitative imaging in medicine and surgery · 2026
- Evaluation of Bone Invasion in Gingival Squamous Cell Carcinoma Using PET/CT Radiomics.Diagnostics (Basel, Switzerland) · 2026
- Photon-counting detector CT with virtual monoenergetic imaging and iterative metal artifact reduction for spinal fusion implants: optimal keV and combined benefits.European journal of radiology · 2026
- Accuracy of CBCT and dental-MRI for detecting peri-implant bone defects at single titanium implants: an in-vitro pilot study.International journal of implant dentistry · 2026
- Radiomics analysis of dynamic carbon-11 methionine PET/computed tomography for differentiating benign and malignant musculoskeletal lesions: a preliminary clinical assessment.Nuclear medicine communications · 2026