Evaluation of virtual monochromatic imaging combined with metal artifact reduction in low-dose dual-energy computed tomography: a phantom study.
Tingting Qu, Ganglian Fan, Yanan Li, Lihong Chen, Le Cao, Jianying Li, Jianxin Guo
Department of Radiology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, China. · GE Healthcare, Computed Tomography Research Center, Beijing, China.
Key points
From the paper's conclusion- MAR consistently improved image quality across titanium-rod positions. In terms of artifact reduction, implant depiction, and adjacent-structure visibility, 60 to 80-keV VM images with MAR provided the most favorable overall performance under the tested low-dose conditions.
Abstract
BACKGROUND: Metal implants can cause substantial artifacts on computed tomography (CT), and the optimal virtual monochromatic (VM) energy for metal artifact reduction (MAR) remains uncertain. This study assessed metal-artifact reduction and overall image quality on low-dose dual-energy computed tomography (DECT) VM images reconstructed with Smart MAR. METHODS: A multi-energy CT phantom (GPH-75B; Kyoto Kagaku Co., Ltd., Kyoto, Japan) was scanned in DECT mode with a 12-mm titanium rod positioned at the center, right, and superior portions. VM images from 40 to 140 keV were reconstructed with MAR and without MAR (no-MAR). Image quality was assessed according to artifact and structure visibility scores, CT values, standard deviations (SDs), titanium-rod dimensions, structural similarity index (SSIM), peak signal-to-noise ratio (PSNR), and root mean square error (RMSE). MAR and no-MAR measurements were compared via two-sided paired t-tests. RESULTS: Across the three titanium-rod positions, MAR generally reduced artifacts, improved titanium-rod edge visibility, and reduced geometric distortion compared with no-MAR. At every rod position and energy level, MAR yielded higher SSIM and PSNR and lower RMSE compared to no-MAR (all P values <0.05). At the central position, the transverse-to-longitudinal diameter ratio was 0.94-1.03 with MAR and 0.21-0.81 with no-MAR (all P values <0.001), remaining close to 1.0 at energies ≥60 keV with MAR. For MAR images, adjacent-structure clarity was highest at 60 keV. In both groups, SSIM and PSNR increased with increasing energy, whereas RMSE and image noise decreased. CONCLUSIONS: MAR consistently improved image quality across titanium-rod positions. In terms of artifact reduction, implant depiction, and adjacent-structure visibility, 60 to 80-keV VM images with MAR provided the most favorable overall performance under the tested low-dose conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Photon-counting detector CT with virtual monoenergetic imaging and iterative metal artifact reduction for spinal fusion implants: optimal keV and combined benefits.European journal of radiology · 2026
- Feasibility of reducing radiation dose and contrast dose while improving image quality in dual-energy CT pulmonary angiography : Use of low-energy virtual monochromatic images with deep learning image reconstruction algorithm.Radiologie (Heidelberg, Germany) · 2026
- Image-Domain CT Metal Artifact Reduction Using an Endpoint-Conditioned Diffusion Bridge.ArXiv · 2026
- Metal Artifact Reduction in Oral Cavity MDCT Using Dental Overlays and the O-MAR Algorithm.Diagnostics (Basel, Switzerland) · 2026
- Photon-Counting Detector CT for Imaging Around Orthopedic and Spinal Metallic Implants: A Systematic Review.Cureus · 2026
- PDDUNet: A primal-dual deep unrolling network for CT metal artifact reduction.Journal of X-ray science and technology · 2026