Off-label Use of the Tandem t: slim X2™ with Control IQ® in Pregnancy: Insights from Five Case Studies.
Ana Rita Leite, Rita Lopes, Patrícia Ferreira, Gabriela Namora, Cristina Gamboa, Joana Queirós, João Sérgio Neves, Sandra Belo
Department of Endocrinology, Diabetes and Metabolism, São João Local Health Unit, Porto, Portugal. · Faculty of Medicine, University of Porto, Portugal.
Key points
From the paper's conclusion- This case series suggests that Tandem t: slim X2™ Control-IQ® is safe and may improve glycemic control during pregnancy and the peripartum period. Further studies comparing this approach with standard care are needed to comprehensively evaluate its effects on glycemic control and maternal and neonatal outcomes.
Abstract
INTRODUCTION/BACKGROUND: Hybrid closed-loop systems (HCLS), such as the Tandem t: slim X2™, are widely used in patients with type 1 diabetes mellitus (T1DM) and have well-established benefits in glycemic control. However, evidence supporting their use during pregnancy and the peripartum period remains limited. CASE PRESENTATION: We conducted a descriptive observational case series from August 2024 to February 2026 of five pregnant women with T1DM using the Tandem t: slim X2™ with Control- IQ® off-label during pregnancy, labor, and the early postpartum period. Glycemic control improved during pregnancy, with mean TIRp increasing from 50.9±17.9% before HCLS initiation to 61.0±14.4% in the third trimester, alongside a reduction in HbA1c from 7.2±1.0% to 6.0±0.5%. Preeclampsia occurred in three of the five cases. In Cases 1 to 4, HCLS remained in automated mode during labor and delivery, with Sleep Activity activated for 33%-40% of the time and latethird- trimester settings maintained. Case 5 suspended HCLS before labor and restarted two days postpartum. Mean pregnancy-specific time in range (TIRp) during labor and delivery was 64.3±10.7%, with minimal hypoglycemia (1.5±2.6%). Early postpartum period (days 1-3) mean TIR was 80.8±14.6%, and time below range (TBR) was 2.3±1.1%. Neonatal hypoglycemias were reported in three of the five cases. CONCLUSION: This case series suggests that Tandem t: slim X2™ Control-IQ® is safe and may improve glycemic control during pregnancy and the peripartum period. Further studies comparing this approach with standard care are needed to comprehensively evaluate its effects on glycemic control and maternal and neonatal outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pregnancy and glycemic outcomes with hybrid closed-loop insulin delivery versus sensor-augmented pump therapy among women with type 1 diabetes.Pregnancy (Hoboken, N.J.) · 2026
- Intrapartum and Early Postpartum Use of Automated Insulin Delivery in Type 1 Diabetes: A Prespecified Analysis of the CIRCUIT Randomized Controlled TrialDiabetes Care · 2026
- Automated Insulin Delivery in Pregnancies Complicated by Type 1 Diabetes: A Real World Observational Study of Available SystemsAustralian and New Zealand Journal of Obstetrics and Gynaecology · 2026
- Closed-Loop Insulin Delivery in Type 1 Diabetes in Pregnancy: The CIRCUIT Randomized Clinical Trial.JAMA · 2025
- Real-world use of automatic insulin delivery pumps in early pregnancy among women living with type 1 diabetes.Diabetic medicine : a journal of the British Diabetic Association · 2026
- Intrapartum Maternal Glycemic Trajectories and the Risk of Neonatal Hypoglycemia in Type 1 Diabetes Pregnancies: A Granular Continuous Glucose Monitoring Analysis.Endocrine practice : official journal of the American College of Endocrinology and the American Association of Clinical Endocrinologists · 2026