K Pump-Mackenzie: Technological Development and Initial Bench Validation of a Fully Implantable 5G-Enabled Ventricular Assist Device.
Luiz Fernando Kubrusly, Osvaldo Malafaia, Benedito Guimarães Aguiar Neto, Paulo Batista Lopes, Roberto Coutinho, Cristiano Akamine, Pollyana Coelho da Silva Notargiacomo, Giovanna Santana,
Faculdade Evangélica Mackenzie do Paraná Curitiba Paraná Brazil Programa de Pós-Graduação em Princípios da Cirurgia, Faculdade Evangélica Mackenzie do Paraná, Curitiba, Paraná, Brazil. · Instituto Presbiteriano Mackenzie São Paulo São Paulo Brazil Instituto Presbiteriano Mackenzie, São Paulo, São Paulo, Brazil.
Key points
From the paper's conclusion- K Pump-Mackenzie demonstrated technical feasibility and satisfactory performance in initial bench validation. The planned preclinical stage will be fundamental to evaluate safety and biological performance, constituting a necessary step for future clinical application, and will take place in 2026.
Abstract
INTRODUCTION: Advanced heart failure is an important cause of mortality in Brazil. Limited access to heart transplantation and the high cost of imported devices reinforce the need for national technological solutions. The K Pump-Mackenzie project proposes the development of a fully implantable mechanical circulatory assist device, integrated with 5G telemetry, with potential application in the unified health system of Brazil, the Sistema Único de Saúde. METHODS: A centrifugal prototype rotor, with real-time control system and integrated flow and pressure sensors, was developed. The device was submitted to laboratory tests to evaluate mechanical performance, operational stability, structural integrity, and connectivity via 5G network. The preclinical protocol in the sheep model was designed and approved by the Comissão de Ética no Uso de Animais, from the Faculdade Evangélica Mackezie do Paraná, in accordance with Conselho Nacional de Segurança Alimentar e Nutricional guidelines. RESULTS: In bench tests, the device showed controlled rotation of up to 10,000 rpm, adjustable flow between 2 and 6 L/min, and average latency of < 20 ms in data transmission. Continuous operational stability was observed for up to 72 hours, with adequate functioning of the monitoring and safety systems. CONCLUSION: K Pump-Mackenzie demonstrated technical feasibility and satisfactory performance in initial bench validation. The planned preclinical stage will be fundamental to evaluate safety and biological performance, constituting a necessary step for future clinical application, and will take place in 2026.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Short-term safety, feasibility, and performance of a novel soft robotic biventricular mechanical circulatory support device avoiding blood contact - The MISSION reBEAT trial.European journal of heart failure · 2026
- A novel magnetically levitated intrapericardial left ventricular assist system in advanced heart failure—2-year results from a prospective, multi-centre studyEuropean Journal of Cardio-Thoracic Surgery · 2026
- A Novel Extracorporeal Membrane Oxygenation System With Enhanced Hemocompatibility and Long-Term Stability.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- The Early Assessment of a Newly Developed Maglev Extracorporeal Mechanical Oxygenation System: 14 Day In Vivo Study Report.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- 5G Cloud-Enabled Telemedicine for Remote Management of Heart Failure Patients With Implantable Cardioverter-Defibrillators: A Multicenter Open-Label Randomized Controlled Trial.MedComm · 2026
- Fontan pump: Advanced preclinical development.The Journal of thoracic and cardiovascular surgery · 2026