Evaluation of Multi-Needle Mesotherapy Needles: Structural Integrity, Solution Uniformity, and Nociceptive Response.
Qinyu Guan, Xin Li, Qinshuang Ou, Ruiyan Wang, Yang Su, Zeyu Fu, Xiaoling Qi
Bloomage Biotech Co., Ltd., Jinan, Shandong, People's Republic of China.
Key points
From the paper's conclusion- While both nine-needle devices maintained structural integrity after repeated use, the test sample demonstrated favorable solution uniformity and lower nociceptive response. Optimizing needle design and manufacturing consistency may improve patient comfort and treatment reliability in mesotherapy.
Abstract
BACKGROUND: Mesotherapy using nine-needle systems has gained popularity in aesthetic medicine, yet key performance parameters-structural integrity, tip alignment, solution uniformity, and nociceptive-remain insufficiently characterized. METHODS: Needle integrity and tip alignment were assessed by scanning electron microscopy (SEM) and industrial computed tomography (CT) before and after 140 porcine-skin insertions. Inter-needle solution uniformity was quantified using a custom 3D-printed collection device. Nociceptive responses were evaluated by electromyographic (EMG) recording of the flexor reflex in three New Zealand White rabbits (n = 3). In each rabbit, the test sample was evaluated on the right foot and the control sample on the left foot. For both the test and control samples, either unused needles or needles that had undergone 140 consecutive punctures in porcine skin were used for the flexor-reflex test to simulate repeated clinical use and assess its effect on the nociceptive response. RESULTS: Both devices maintained structural continuity after 140 insertions, with no fracture. SEM revealed that control samples exhibited tip bending, blunting, edge irregularity, and increased surface roughness, whereas test samples retained a sharp, symmetric tip with minimal morphological change. The test samples showed significantly lower exit errors than the control group both before insertion (±0.0173-0.0240 vs ±0.0315-0.0432, P≤0.01) and after insertion (±0.0155-0.0242 vs ±0.0322-0.0431, P≤0.01). In comparison, the alignment of the test samples was more consistent (the detection error of the instrument was controlled within ±0.05 mm). Solution uniformity analysis showed a narrower deviation range for the test sample, indicating improved flow consistency. EMG showed that the test sample induced lower flexor reflex amplitudes than the control, both before (608-879 μV vs 462-1458 μV) and after 140 insertions (68-425 μV vs 234-855 μV), indicating reduced nociceptive stimulation. CONCLUSION: While both nine-needle devices maintained structural integrity after repeated use, the test sample demonstrated favorable solution uniformity and lower nociceptive response. Optimizing needle design and manufacturing consistency may improve patient comfort and treatment reliability in mesotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Scanning Electron Microscopic Evaluation of Bevel Deformation in Dental Needles After Repeated Anesthetic InsertionsJournal of Maxillofacial and Oral Surgery · 2026
- Influence of Local Anesthesia Techniques on the Microstructural Integrity of Dental Injection Needles: A Field Emission Scanning Electron Microscopy Analysis.Bioengineering (Basel, Switzerland) · 2026
- Multimethod Evaluation of the Novel Reciproc Minima System: Geometric Design, Mechanical Performance, and Irrigation Dynamics.Dentistry journal · 2026
- High-resolution spatiotemporal mapping of transient Joule heating and skin tissue responses during microneedle electroporation.Bioelectrochemistry (Amsterdam, Netherlands) · 2026
- Discharge Mode Enhances the Energy Efficiency of Radiofrequency Microneedle: An In Vivo Porcine Study of Optimized Histological Responses Under Constant Total Energy.Journal of cosmetic dermatology · 2026
- Biomechanical response and post-intervention pain after dry needling, percutaneous electrolysis, and percutaneous electrical nerve stimulation on upper trapezius muscle.Journal of tissue viability · 2026