Medical devices (Auckland, N.Z.)

Improved Hemostatic Performance of the Lexington Medical AEON Endoscopic Stapler Narrow Reload Compared with the Medtronic Signia Small Diameter Reload in a Canine Pulmonary Lobectomy Model.

Dennis Begos, Carolyn M Mothersell, Dan Smeak

Lexington Medical, Bedford, MA, USA. · Department of Veterinary Clinical Sciences, Colorado State University College of Veterinary Medicine and Biomedical Sciences, Fort Collins, CO, USA.

Key points

From the paper's conclusion
  • The Lexington Medical AEON Endoscopic Stapler Narrow Reload showed superior hemostatic results on pulmonary vessels in a canine model compared with the Medtronic Signia Small Diameter Reloads.

Abstract

PURPOSE: To evaluate stapling performance of the Lexington Medical AEON Endoscopic Powered Stapler Handle and AEON Endoscopic Stapler Narrow Reload (Test Reloads, TR) compared to the currently available Medtronic Signia Small Diameter Reloads (Control Reloads, CR). This report evaluates acute surgical results for hemostasis, stapler handling, and performance. PATIENTS AND METHODS: Eight (8) female canines (Mongrel, 15-25 kg) were used. Five animals were test animals, and three were controls. Each animal underwent pulmonary lobectomy using a stapler with either TR or CR fitted to the appropriate handle. At least two white and gray cartridges were used for the pulmonary vessels in each animal. A Likert scale of 1-5 was used to assess staple line bleeding in real time. The vessel size and mean arterial pressure (MAP) were recorded at the time of firing. Animals were euthanized on day 21 (± 2 days) for histological analysis of the staple sites (data not included in this report). RESULTS: A total of 29 firings were performed in the test animals (mean 5.8/animal) and 19 in the control animals (mean 6.3/animal). The size of the vessels ranged from 3-9.5 mm in the test group, from 3-7.5 mm in the control group. The Likert scale for all firings of the AEON stapler was 1 (no bleeding at the tissue site after the initial blotting of the staple line). The Likert scale average for the Signia stapler was 1.8, due to four scores of 5 (significant bleeding requiring intervention, such as extensive coagulation or ligation with clips). This difference was statistically significant (p = 0.017, Mann-Whitney U-test comparing per-animal mean scores). The mean vessel size (p=0.287) and MAP (p=0.518) did not differ between the 2 groups. CONCLUSION: The Lexington Medical AEON Endoscopic Stapler Narrow Reload showed superior hemostatic results on pulmonary vessels in a canine model compared with the Medtronic Signia Small Diameter Reloads.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Improved Hemostatic Performance of the Lexington Medical AEON Endoscopic Stapler Narrow Reload Compared with the Medtronic Signia Small Diameter Reload in a Canine Pulmonary Lobectomy Model — Science Skim