Key points
From the paper's conclusion- Navigation was associated with a markedly lower rate of acute symptomatic implant malposition during lateral or PL transiliac placement of a PTI.
Abstract
OBJECTIVE: To determine whether use of intraoperative navigation is associated with a reduced rate of acute symptomatic implant malposition during iliosacral implant placement in sacroiliac (SI) joint fusion and sacral fracture fixation. METHODS: Retrospective analysis of post-market databases regarding device use and reported device-related complaints from a single device manufacturer between March 2021 and August 2026 were combined, including procedure type, procedure date and use of navigation systems. Analysis focused on lateral and posterolateral (PL) transiliac placement of a porous threaded implant (PTI, iFuse TORQ) during SI joint fusion and sacral fracture fixation procedures performed in the USA. All acute symptomatic implant malpositions reported as product complaints were identified and linked to index surgeries through manual identification. The proportion of navigated and non-navigated cases that resulted in symptomatic implant malposition was compared using Firth's penalized logistic regression test. RESULTS: 25,096 procedures using 68,352 PTI were identified. Navigation was used in 15.6% (n=3,919) of cases. Implant malposition causing acute symptoms occurred in 63 cases (overall rate 0.25%). Of these, 2 occurred in navigated cases (rate, 0.05%) and 61 occurred in non-navigated cases (rate, 0.29%, odds ratio 0.22, 95% CI 0.046-0.639, p=0.0014). CONCLUSION: Navigation was associated with a markedly lower rate of acute symptomatic implant malposition during lateral or PL transiliac placement of a PTI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy, complications, and outcomes of dual-pin reference frame placement in navigated minimally invasive transforaminal lumbar interbody fusion.Journal of neurosurgery. Spine · 2026
- Intraoperative Accuracy of Sacroiliac Joint Fusion Implant Placement Using 3D Computed Tomography-Guided Imaging for Primary Sacroiliac Arthropathy.International journal of spine surgery · 2026
- A novel pinless augmented reality-based navigation system for total hip arthroplasty in the supine position: a comparative study of acetabular cup angle accuracyArchives of Orthopaedic and Trauma Surgery · 2026
- Image guidance is associated with lower rates of pseudoarthrosis diagnosis after long-segment posterior spinal instrumentation: a propensity-matched cohort study.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026
- Accuracy of 3D-navigated screw fixation in pelvic ring fractures: a single-centre consecutive observational case seriesEuropean Journal of Orthopaedic Surgery & Traumatology · 2025
- Robotic navigation-assisted versus conventional freehand placement of the infra-acetabular screw in acetabular fracture surgery: A retrospective comparative study.Injury · 2026