A Novel Device for Pulmonary Expansion Under Simulated Healthy and Neuromuscular Conditions: A Bench Simulation Study.
Alessandra Carneiro Dorça, Letícia de Araújo Morais
Alessandra Dorça Institute, Goiânia, Goiás, Brazil. · Undergraduate Program in Physiotherapy, Estácio de Sá University Center in Goiás, Goiânia, Goiás, Brazil.
Key points
From the paper's conclusion- Pressure and volume were higher with the new device in both simulated models. Flow did not differ significantly among devices, although lower variability was descriptively observed with the new device.
Abstract
BACKGROUND: Amyotrophic Lateral Sclerosis (ALS) leads to a progressive decline in respiratory function, requiring volumetric expansion techniques to prevent hypoventilation. However, the mechanical performance of devices used for this purpose requires evaluation under controlled conditions in bench simulation studies. Devices, such as the manual resuscitator, are commonly used, but present limitations, including high pressures, variable flow, among others. A novel respiratory muscle training device designed to provide controlled flow and pressure (ILLIFE-T) was developed. OBJECTIVE: To evaluate the behavior of the variables pressure, volume, and flow during the volumetric expansion technique, comparing the new device with the manual resuscitator and the manual resuscitator + Positive Expiratory Pressure (PEP), using simulated models of healthy and ALS affected lungs. METHODS: This bench simulation study used the ASL 5000 simulator, configured with predefined respiratory profiles representing healthy and ALS respiratory conditions. Four repetitions of the volumetric expansion maneuver were performed for each device and simulated respiratory condition, with five insufflations per repetition. One recording was randomly selected from each repetition for analysis, resulting in four observations per device and simulated respiratory condition. RESULTS: In the simulation with healthy respiratory conditions, the new device generated higher pressure (23.38±3.81 cmH2O), volume (1991.25 ±174.10 mL), and lower flow variability (-1.47±11.43 L/min) compared with the manual resuscitator (12.25±11.73 cmH2O; 1113.64±131.75 mL; 18.84±61.31 L/min) and the manual resuscitator with PEP valve (16.89±0.64 cmH2O; 1817.25±86.77 mL; -49.49±9.96 L/min), respectively. In ALS simulated models, the new device also achieved higher pressure (23.69±3.29 cmH2O) and volume (2235.28±264.74 mL) compared with the manual resuscitator (8.33±8.34 cmH2O; 563.45±212.03 mL, respectively) and the manual resuscitator with PEP valve (14.89±7.60 cmH2O; 1449.86±153.09 mL, respectively). Flow values were similar among devices: 15.10 ± 18.00 L/min for the new device, 19.28±56.03 L/min for the manual resuscitator, and 0.18±51.67 L/min for the manual resuscitator with PEP valve. CONCLUSION: Pressure and volume were higher with the new device in both simulated models. Flow did not differ significantly among devices, although lower variability was descriptively observed with the new device.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Manikin-based noninferiority simulation study of Airgency™: A novel automatic bag valve mask device for emergency respiratory support.Turkish journal of emergency medicine · 2026
- VENTILA2: A Fuzzy Logic-Based Simulation and Decision-Support Framework for Pressure-Controlled Ventilation-A Proof of Concept.Healthcare (Basel, Switzerland) · 2026
- Beyond the Larynx: New Insights Into Upper Airway Responses to Mechanical Insufflation-Exsufflation.Respirology case reports · 2026
- Improved Suction Efficiency and User Satisfaction With a Novel Portable Manual Suction Device: A Simulation-Based Pre-Hospital Study.Inquiry : a journal of medical care organization, provision and financing · 2026
- Comparison of bag-valve-device ventilation and volume-controlled ventilation with laryngeal tube evo in intra-arrest ventilation-a prospective randomized study in human body donors.Journal of anesthesia, analgesia and critical care · 2026
- Evaluation of aerosol delivery in ventilated patients via vibrating mesh nebuliser versus other delivery devices using a patient-specific in silico whole-lung model.Respiratory research · 2026