A Novel Device for Pulmonary Expansion Under Simulated Healthy and Neuromuscular Conditions: A Bench Simulation Study.

Alessandra Carneiro Dorça, Letícia de Araújo Morais

Alessandra Dorça Institute, Goiânia, Goiás, Brazil. · Undergraduate Program in Physiotherapy, Estácio de Sá University Center in Goiás, Goiânia, Goiás, Brazil.

Computational

Key points

From the paper's conclusion
  • Pressure and volume were higher with the new device in both simulated models. Flow did not differ significantly among devices, although lower variability was descriptively observed with the new device.

Abstract

BACKGROUND: Amyotrophic Lateral Sclerosis (ALS) leads to a progressive decline in respiratory function, requiring volumetric expansion techniques to prevent hypoventilation. However, the mechanical performance of devices used for this purpose requires evaluation under controlled conditions in bench simulation studies. Devices, such as the manual resuscitator, are commonly used, but present limitations, including high pressures, variable flow, among others. A novel respiratory muscle training device designed to provide controlled flow and pressure (ILLIFE-T) was developed. OBJECTIVE: To evaluate the behavior of the variables pressure, volume, and flow during the volumetric expansion technique, comparing the new device with the manual resuscitator and the manual resuscitator + Positive Expiratory Pressure (PEP), using simulated models of healthy and ALS affected lungs. METHODS: This bench simulation study used the ASL 5000 simulator, configured with predefined respiratory profiles representing healthy and ALS respiratory conditions. Four repetitions of the volumetric expansion maneuver were performed for each device and simulated respiratory condition, with five insufflations per repetition. One recording was randomly selected from each repetition for analysis, resulting in four observations per device and simulated respiratory condition. RESULTS: In the simulation with healthy respiratory conditions, the new device generated higher pressure (23.38±3.81 cmH2O), volume (1991.25 ±174.10 mL), and lower flow variability (-1.47±11.43 L/min) compared with the manual resuscitator (12.25±11.73 cmH2O; 1113.64±131.75 mL; 18.84±61.31 L/min) and the manual resuscitator with PEP valve (16.89±0.64 cmH2O; 1817.25±86.77 mL; -49.49±9.96 L/min), respectively. In ALS simulated models, the new device also achieved higher pressure (23.69±3.29 cmH2O) and volume (2235.28±264.74 mL) compared with the manual resuscitator (8.33±8.34 cmH2O; 563.45±212.03 mL, respectively) and the manual resuscitator with PEP valve (14.89±7.60 cmH2O; 1449.86±153.09 mL, respectively). Flow values were similar among devices: 15.10 ± 18.00 L/min for the new device, 19.28±56.03 L/min for the manual resuscitator, and 0.18±51.67 L/min for the manual resuscitator with PEP valve. CONCLUSION: Pressure and volume were higher with the new device in both simulated models. Flow did not differ significantly among devices, although lower variability was descriptively observed with the new device.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A Novel Device for Pulmonary Expansion Under Simulated Healthy and Neuromuscular Conditions: A Bench Simulation Study — Science Skim