Microscope-Integrated Quantitative Fluorescence Imaging for Assessing Crack Severity in Teeth.
Hye-Min Ku
Department of Conservative Dentistry, Oral Science Research Center, Yonsei University College of Dentistry, Seoul 03722, Korea.
Key points
From the paper's conclusion- The microscope-QLF hybrid technique showed a measurable and consistent relationship with QLF-D for quantitative fluorescence assessment. Although systematic differences were observed, the hybrid technique may serve as a supportive tool for assessing relative crack severity during endodontic procedures.
Abstract
BACKGROUND: This study investigated the fluorescence characteristics associated with cracks and evaluated the relationship between crack depth and fluorescence-derived variables using quantitative light-induced fluorescence-digital (QLF-D) and microscope-QLF hybrid techniques. Agreement between maximum fluorescence loss (ΔFmax) values obtained from the two modalities was also assessed. METHODS: Extracted teeth with confirmed cracks were examined using QLF-D and a microscope-QLF hybrid technique integrating a fluorescence light source with an operating microscope. Fluorescence-derived variables, including ΔFmax and red fluorescence (ΔR) and maximum red fluorescece (ΔRmax) variables, were quantified using dedicated analysis software. The crack depth was measured using micro-computed tomography as the reference standard. Associations between the crack depth and fluorescence-derived variables were analyzed using Spearman's rank correlation analysis. Agreement between ΔFmax values obtained from the two imaging modalities was evaluated using Bland-Altman analysis. RESULTS: Cracks displayed characteristic fluorescence loss and red fluorescence patterns in both imaging modalities. Among the fluorescence-derived variables, ΔFmax demonstrated the strongest correlation with crack depth for both the QLF-D and the microscope-QLF hybrid techniques (p<0.01). The corresponding ΔFmax values obtained from the two modalities were strongly correlated. Bland-Altman analysis indicated a small systematic bias, with the microscope-QLF hybrid technique slightly overestimating ΔFmax; however, all measurements remained within the 95% limits of agreement. CONCLUSION: The microscope-QLF hybrid technique showed a measurable and consistent relationship with QLF-D for quantitative fluorescence assessment. Although systematic differences were observed, the hybrid technique may serve as a supportive tool for assessing relative crack severity during endodontic procedures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optical-Fluorescence Decomposition Reveals Progressive Multimodal Divergence with Histologic Severity of Occlusal Caries.Photodiagnosis and photodynamic therapy · 2026
- Paired in vivo diagnostic accuracy and reliability of QrayCam Pro and QrayPen for occlusal and proximal caries in permanent posterior teeth.BMC oral health · 2026
- Evaluation of 18-Month Clinical Performance of CAD/CAM Hybrid Onlay Restorations Using FDI Criteria and QLF Imaging.International journal of computerized dentistry · 2026
- Fluorescence compatibility of various resin composites on cervical restorations: An in-vitro evaluation.Clinical oral investigations · 2026
- In Vivo Comparison of Laser Fluorescence, Near-Infrared Transillumination, and Conventional Methods for Approximal Caries Detection.Journal of clinical medicine · 2026
- Effects of imaging conditions on fluorescence intensity during fluorescence-aided root canal treatment.Photodiagnosis and photodynamic therapy · 2026