Accuracy of tympanometry cutoff points in children from Central-Western Brazil.
Cláudia Cruvinel Câmara, Melissa Ameloti Gomes Avelino, Paulo Sergio Sucasas da Costa, Alexandre Siqueira Guedes Coelho, Maria Claret Costa Monteiro Hadler
Programa de Pós-graduação em Ciências da Saúde, Faculdade de Medicina, Universidade Federal de Goiás - UFG - Goiânia (GO), Brasil. · Programa de Pós-graduação em Nutrição e Saúde, Universidade Federal de Goiás - UFG - Goiânia (GO), Brasil.
Key points
From the paper's conclusion- The cutoff point of -180 daPa combined with static compliance <0.2 mL improved screening accuracy and reduced unnecessary referrals. It provided the best balance between specificity, sensitivity, and accuracy, aligning more closely with TEOAEs results and supporting updating this tympanometric criterion in infant hearing screening.
Abstract
PURPOSE: This study evaluated tympanometric cutoff points (tympanometric peak pressure and static compliance) for transient evoked otoacoustic emissions (TEOAEs) and determined their diagnostic accuracy. METHODS: This analytical study, based on a randomized clinical trial, included 84 children (6-42 months; median 24 months) from public childcare centers in Brazil. Associations between cutoff points and TEOAEs were verified by Fisher's exact test and Chi-square. Kappa assessed agreement between ears, and the correlation between tympanometric peak pressure and static compliance was analyzed by Spearman. Diagnostic accuracy was determined using ROC (Receiver Operating Characteristic) curves and confirmed with EpiDat 3.1. RESULTS: In the 84 children, 66 right ears and 73 left ears were examined; however, 42 ears were randomly selected from each side for the final analysis. A significant correlation was found between peak pressure and static compliance. ROC analysis identified -180 daPa as the most accurate cutoff (AUC = 0.781; 95% CI: 0.656-0.905; p < 0.001). EpiDat indicated a specificity of 93.9%, a sensitivity of 55.6%, and an overall accuracy of 85.7%. Static compliance <0.2 mL showed comparable performance (specificity = 98.5%, sensitivity = 50.0%, accuracy = 88.1%). CONCLUSION: The cutoff point of -180 daPa combined with static compliance <0.2 mL improved screening accuracy and reduced unnecessary referrals. It provided the best balance between specificity, sensitivity, and accuracy, aligning more closely with TEOAEs results and supporting updating this tympanometric criterion in infant hearing screening.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Application of wideband tympanometry combined with kempthorne criteria in the diagnosis of otitis media with effusion in adults: a prospective study.Frontiers in surgery · 2026
- Otoscopic and Audiometric Clues for Suspecting OMAAV in Elderly Patients With Refractory Otitis Media With Effusion.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Initial High-Frequency Amplification Accuracy Predicts Long-Term Hearing Aid Fitting Outcomes: A Real Ear Measurement-Based Study.Journal of clinical medicine · 2026
- Estimating the natural resolution of otitis media with effusion and associated hearing outcomes in a paediatric population in Western Australia: a retrospective analysis.International journal of pediatric otorhinolaryngology · 2026
- Preoperative Tympanometry and the Presence of Middle Ear Effusion at the Time of Tympanostomy Tube PlacementLaryngoscope Investigative Otolaryngology · 2026
- Screening Utility of Ambient Pressure Tympanometry for the Detection of Radiologic Correlates of Subjective Pulsatile Tinnitus.Otology & neurotology open · 2026