Comparison between two methods of cleaning and disinfection of digital pulse oximeters in primary care.
Juliana Arruda de Matos, Ivana Bogado Martinez, Érika de Almeida Leite da Silva Teixeira de Souza, Hilda Silva de Souza Barreto, Maria Cristina da Silva Lourenço, Regina Paiva Daumas
Oswaldo Cruz Foundation, National Institute of Infectology Evandro Chagas, Immunization and Health Surveillance Research Unit. Rio de Janeiro, Rio de Janeiro, Brazil. · Oswaldo Cruz Foundation, National Institute of Infectology Evandro Chagas, Bacteriology and Bioassays Laboratory. Rio de Janeiro, Rio de Janeiro, Brazil.
Key points
From the paper's conclusion- the efficacy of the two methods was similar and suboptimal, highlighting the challenges associated with the cleaning and disinfection of electronic medical devices in primary care.
Abstract
OBJECTIVES: to compare the efficacy and duration of cleaning and disinfection of digital oximeters with two methods in a primary care setting. METHODS: microbiological culture and adenosine triphosphate assay results were evaluated in oximeters inoculated with Salmonella typhimurium, Staphylococcus aureus or Escherichia coli: after water and detergent followed by 70% ethanol (method 1); after one-step disinfectant based on benzalkonium chloride and polyhexamethylene biguanide hydrochloride (method 2); and no cleaning and disinfection (control). Effective disinfection was defined as: (1) microbial load reduction of ≥99.99%, (2) negative culture or (3) final microbial load <103. RESULTS: effective disinfection varied between 33.3% (criterion 2) and 72.2% (criterion 3) and was similar between the two methods, while cleaning and disinfection duration in method 2 was notably shorter. CONCLUSIONS: the efficacy of the two methods was similar and suboptimal, highlighting the challenges associated with the cleaning and disinfection of electronic medical devices in primary care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Contingency Methods for Cleaning Single-Use Medical Equipment in Crisis, Disaster, or War.Prehospital and disaster medicine · 2026
- A-396 Beyond the “Meter Disinfected” Button: A Proactive Quarterly Audit Approach to Glucometer Disinfection For Patient SafetyClinical Chemistry · 2026
- A PBMC-based assay detects residual pro-inflammatory response in endoscopes after high-level disinfection undetected by ATP bioluminescence: a pilot study.The Journal of hospital infection · 2026
- Evaluating the efficacy of disinfecting agents against Staphylococcus aureus on external fixator devices.Injury · 2026
- Chlorhexidine versus povidone-iodine for surgical site infection prevention: an updated meta-analysis and trial sequential analysis of randomized controlled trialsFrontiers in Medicine · 2025
- Microbial contamination and decontamination of reusable digital intraoral imaging equipment: a narrative review.Frontiers in oral health · 2026