Microbial barrier integrity of 3-way-stopcocks following disinfection with and without mechanical challenge.
Tobias Hans Conrad, Hagen Frickmann, Andreas Podbielski, Philipp Warnke
1Institute for Medical Microbiology, Virology and Hygiene, University Medicine Rostock, Rostock, Germany.
Key points
From the paper's conclusion- The 3-way stopcocks showed a robust barrier microbial integrity against long-term disinfection without and with combined mechanical challenging. The results suggest that infection prevention and control risks due to stress cracks caused by disinfection of 3-way stopcocks seems to be negligible and that their disinfection is safe to be performed.
Abstract
BACKGROUND: The study assessed the microbial barrier integrity of 3-way stopcocks exposed to various disinfecting agents in terms of stress cracks with potential relevance for infection prevention and control. METHODS: Four different 3-way stopcocks were filled with medium containing each more than 100,000 colony-forming units of Staphylococcus epidermidis, Pseudomonas aeruginosa, Escherichia coli and Candida albicans, with subsequent exposure to four different disinfecting agents for 6 and 24 h as well as for 28 days, exceeding realistic real-word exposure times. Mechanical challenge was simulated by repeated docking of syringes, combined with 24 h disinfection turns. After the disinfection terms, lacking bacterial growth from closed 3-way stopcocks as well as positive growth from the bacterial suspension inside was confirmed in brain heart infusion broth. RESULTS: Absence of bacterial growth from closed 3-way stopcocks and remaining vitality of the bacteria in the suspension within the stopcocks could be confirmed in all instances. CONCLUSIONS: The 3-way stopcocks showed a robust barrier microbial integrity against long-term disinfection without and with combined mechanical challenging. The results suggest that infection prevention and control risks due to stress cracks caused by disinfection of 3-way stopcocks seems to be negligible and that their disinfection is safe to be performed.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating the efficacy of disinfecting agents against Staphylococcus aureus on external fixator devices.Injury · 2026
- In-Situ Disinfection of Hospital Privacy Curtains Using a Next-Generation Aerosolized Hydrogen Peroxide SystemmedRxiv · 2026
- In Vitro Inhibitory Activity of Three Hospital Biocides Against Multidrug-Resistant Clinical Isolates: A Pilot Study From Morocco.Cureus · 2026
- Complementary Time-Kill Activity Displayed by Povidone-Iodine and Hydrogen Peroxide Against Shoulder Arthroplasty-Associated Pathogens, Envisaging Intraoperative Irrigation.Pharmaceuticals (Basel, Switzerland) · 2026
- Triclosan Surface Coating Inhibits Bacterial Attachment: In Vitro Assessment for Prosthetic Joint Infection Prevention.BioMed research international · 2026
- Comparative Evaluation of Microbial Transfer in Intermittent Catheters: An in vitro Study of Protective Features Against Uropathogens.Journal of wound, ostomy, and continence nursing : official publication of The Wound, Ostomy and Continence Nurses Society · 2026