Not all sugars are equal: galactose interference leading to false hyperglycaemia in a case of classical galactosaemia.
Sophie Manoy, Philip Crook, Kevin Gaughan, Roshni Vara, Angharad Flower, Taiwo Oladimeji, Caralyn Jarvis, Adam Pope,
Queensland Lifespan Metabolic Medicine Service, Queensland Children's Hospital, Brisbane, Australia. · Biochemical Genetics, Synnovis Group LLP, London, England.
Key points
From the paper's conclusion- Identification of discrepancies in certain POC glucometers and capillary blood gas or venous glucose measurements by the clinical biochemistry laboratory may indicate POC glucometer interference and raise suspicion for a diagnosis of classical galactosaemia in the newborn setting.
Abstract
OBJECTIVES: Classical galactosaemia (OMIM #230400) is an inborn error of carbohydrate metabolism caused by deficiency of the enzyme galactose-1-phosphate uridyl transferase (GALT). CASE PRESENTATION: We report a case of a female infant who presented with vomiting, failure to thrive, hepatic transaminitis and a discrepancy between point-of-care (POC) glucometer (Accu-chek® Inform II, Roche Diagnostics, Germany) and capillary blood glucose (ABL90 Flex Plus© blood gas analyser) testing results, raising the suspicion for a diagnosis of classical galactosaemia. This was attributed to suspected analytical interference from markedly elevated blood galactose, due to a known limitation of the POC glucometer specificity for the monosaccharide sugars glucose and galactose. This led to falsely elevated glucose readings on POC glucometer testing in the presence of a high galactose concentration. Reduced GALT enzyme activity and genetic testing subsequently confirmed a diagnosis of classical galactosaemia. CONCLUSIONS: Identification of discrepancies in certain POC glucometers and capillary blood gas or venous glucose measurements by the clinical biochemistry laboratory may indicate POC glucometer interference and raise suspicion for a diagnosis of classical galactosaemia in the newborn setting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-156 Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 in for Hemoglobin H Disease ScreeningClinical Chemistry · 2026
- Measurement of galactose-1-phosphate in dried blood spots by LC-MS/MS: method characteristics and therapeutic range.Molecular genetics and metabolism · 2026
- Blood Glucose Monitoring in the Neonatal Period: A Study on Evaluation Using the RAPIDPoint 500/500e.Pediatrics international : official journal of the Japan Pediatric Society · 2026
- Comparison of Blood Glucose Measurement Methods at Low Concentrations in Neonates at Risk for Hypoglycemia.The journal of applied laboratory medicine · 2026
- Prenatal consequences of GALT deficiency in a rat model of classic galactosemia.PLoS genetics · 2026
- B-030 Analytical Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 (High-Performance Liquid Chromatography)Clinical Chemistry · 2026