A-156 Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 in for Hemoglobin H Disease Screening
Jing Liu, Tianqi Yuan, Jian Wan, Hailian Zhou, Yuxiang Qiu, Ting Wu, Qiuyang Guo, Wen Zeng,
Key points
From the paper's conclusion- This study aimed to evaluate the performance of the Maccura G 02 in screening for Hb H disease.
Abstract
Background Hemoglobin H (Hb H) disease is a hereditary hemolytic anemia characterized by the severe deficiency in a-chain synthesis due to a-globin gene deletions or mutations, leading to a relative excess of ß-chains. Due to the high incidence of Hb H disease, which affects the structure, quantity, and glycation capacity of hemoglobin, accurate measurement of glycated hemoglobin (HbA1c) in patients with Hb H disease is critically important. This study aimed to evaluate the performance of the Maccura G 02 in screening for Hb H disease. Methods HbA1c analysis was performed on 20 patients with Hb H disease using the Maccura G 02. These results were compared with Tosoh G11, Bio-Rad D-100, and Mass Spectrometry (MS). Furthermore, genetic testing was conducted to provide definitive confirmation of the patients’ diagnoses. Results The G 02 analyzer displayed a total of 18 alarm messages, all indicating an abnormal peak of HbA1a, with an alarm rate of 90%. Compared to current mainstream high-end analyzers, the G 02 demonstrated a higher alarm rate for identification, while the Tosoh G11 and Bio-Rad D-100 show rate of 36.4% and 77.2%, respectively. Using MS as the reference standard, the concordance rate of the G 02 was 95%. Notably, compared with the Tosoh G11 (70%) and D-100 (100%), the anti-interference qualification rates of the G 02 was comparable to or better than those of current mainstream instruments. Additionally, genetic testing was performed on all 20 patients, and the diagnostic results were -a3.7/--SEA (n=18) as well as --SEA/aCSa (n=2), validating the G 02’s performance in screening for Hb H disease. Conclusion This study demonstrates the excellent performance of the Maccura G 02 in screening for Hb H disease. Its early warning capability and result reliability are comparable to or superior to those of current mainstream glycated hemoglobin analyzers (High-Performance Liquid Chromatography method). Furthermore, its robust capability in identifying Hb H disease extends the performance advantages of Maccura’s glycation product line.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hb J-Guantanamo: a Rare Hemoglobin Variant Identified in Morocco Following Unexpected HbA1c Results.Clinical laboratory · 2026
- Abnormally Low Hemoglobin A1c Due to a Heterozygous Alpha-Globin Variant (Hb I; HBA2 c.49A>G, p.Lys17Glu) in a Geriatric Patient.Cureus · 2026
- B-030 Analytical Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 (High-Performance Liquid Chromatography)Clinical Chemistry · 2026
- Glycated Haemoglobin Versus Serum Fructosamine in Assessing Glycaemic Control Amongst Type 2 Diabetes Mellitus Patients With Haemoglobin Variants: A Systematic Review.BioMed research international · 2026
- B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic VariantsClinical Chemistry · 2026
- B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine MethodsClinical Chemistry · 2026