Clinical chemistry and laboratory medicine

Analytical performance evaluation of the new double-antibody sandwich Mindray total vitamin D assay and comparison vs. liquid chromatography tandem mass spectrometry.

Konstantinos Makris, Ioulia Trifonidi, Pierre Lukas, Efstathios Chronopoulos, Etienne Cavalier

Clinical Biochemistry Department, KAT General Hospital, Kifissia, Athens, Greece. · Department of Clinical Chemistry, University of Liege, Liege, Belgium.

Key points

From the paper's conclusion
  • The Mindray assay showed excellent analytical performance over the evaluated range and matrices. Its initial difference from the LC-MS/MS method was reduced after regression recalibration. The findings emphasize the importance of reference-procedure alignment and verification of method bias at clinically relevant concentrations.

Abstract

OBJECTIVES: Measurement of total 25-hydroxyvitamin-D [25(OH)D] is widely used to assess vitamin D status, but immunoassays may show method-dependent differences related to calibration, analyte recognition, and metabolite effects. The new Mindray 25(OH)D assay is a noncompetitive chemiluminescent immunoassay based on an anti-metatype antibody design. METHODS: We evaluated calibration stability, precision, linearity, recovery, limit of quantification (LoQ), and potential effects of 24,25(OH)2D3, C3-epi-25(OH)D3 and 25(OH)D2. Serum and K3-EDTA plasma were compared in 150 paired samples. Comparison with a VDSP-certified LC-MS/MS was assessed in 146 serum samples. A regression derived recalibration equation was subsequently evaluated in an independent cohort of 115 samples. RESULTS: Total coefficients of variation (CVs) were 2.05-5.69 % across concentrations. Calibration remained within the predefined acceptance criteria for 31 days. The assay was linear from 3.92 to 136 μg/L (R2=0.9998), and the LoQ was 2.78 μg/L (CV=3.65 %). Mean recovery across 33 mixtures was 101.31 % (95 % CI, 100.47-102.15 %), with results ranging from 96.14 to 105.12 %. No clinically relevant effects of the investigated vitamin D metabolites were observed within the studied ranges. Serum and EDTA-plasma results were highly concordant. The CCC of the immunoassay vs. LC-MS/MS was 0.940, with a mean Bland-Altman difference of -2.5451 μg/L; after recalibration in the independent cohort, CCC increased to 0.995 and the mean difference was +0.0045 μg/L. CONCLUSIONS: The Mindray assay showed excellent analytical performance over the evaluated range and matrices. Its initial difference from the LC-MS/MS method was reduced after regression recalibration. The findings emphasize the importance of reference-procedure alignment and verification of method bias at clinically relevant concentrations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers