Comparative assessment of cross-reactivity with Big-prolactin and macroprolactin in five commercial assays for prolactin.
Romain Vankemmel, Sofiane Chaib, Marine Malléa-Hivert, Emilie Merlen, Benoit Soudan, Pascal Pigny
Laboratoire de Biochimie Hormonologie, Centre de Biologie Pathologie, CHU, Lille, France. · Service d'Endocrinologie, Diabétologie, Oncologie Endocrinienne et Métabolisme, Hôpital Claude Huriez, CHU, Lille, France.
Key points
From the paper's conclusion- All five immunoassay platforms recognize the Big-PRL component in serum but to a variable extent. A high proportion of Big-PRL, combined with the use of a highly cross-reactive assay, may cause a false diagnosis of hyperprolactinemia.
Abstract
OBJECTIVES: Cross-reactivity with high molecular-weight (MW) macroprolactin in prolactin (PRL) immunoassays has been the subject of numerous studies. In contrast, little data are available regarding Big-PRL, a 50-60 kDa PRL isoform. METHODS: The immunoreactivity of Big-PRL was evaluated on five platforms (Alinity (Abbott), DxI 800 (Beckman Coulter), Liaison XL (DiaSorin), Kryptor compact plus (Thermo Scientific), and iSYS (IDS Immunodiagnostic Systems)) using 10 serum samples characterized by gel filtration chromatography (GFC) and containing between 17 and 80 % of Big-PRL. The spread of the results as a function of Big-PRL percentage was determined. Bland Altman bias between assays was calculated using a panel of GFC-characterised sera containing macroprolactin (n=46) or monomeric PRL (n=45). RESULTS: The Big-PRL samples were obtained from nonpregnant women, seven of whom had galactorrhea or oligospaniomenorrhea. All assays recognized Big-PRL in undiluted serum but to a variable extent. The spread of total PRL values correlated positively with Big-PRL percentage (r=0.954, p<0.0001). In samples containing >70 % Big-PRL, the methods ranked in the same order of recognition as that observed with macroprolactin-rich samples. Big-PRL was strongly detected by Alinity and iSYS whereas Kryptor and DxI were less sensitive to Big-PRL, as observed with macroprolactin. Consequently, the risk of false diagnosis of hyperprolactinemia tended to be lower with these two last assays. CONCLUSIONS: All five immunoassay platforms recognize the Big-PRL component in serum but to a variable extent. A high proportion of Big-PRL, combined with the use of a highly cross-reactive assay, may cause a false diagnosis of hyperprolactinemia.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishment of Sex-Specific Reference Intervals for Total and Post-PEG Prolactin on an Automated Analyzer.The journal of applied laboratory medicine · 2026
- Optimization of PEG precipitation for macroprolactinemia screening.Clinical chemistry and laboratory medicine · 2026
- 386 False Hook Effect in Prolactin Testing Due to Reagent-Associated Signal Interference on the Siemens Atellica AnalyzerAmerican Journal of Clinical Pathology · 2026
- Macrohormones in endocrinology: prevalence, clinical impact, and diagnostic approaches.Frontiers in endocrinology · 2026
- Serum prolactin as a shared inflammation-associated signal across inflammatory breast conditions: a four-group comparative study with exploratory inflammatory breast cancer subtype analysis.Gland surgery · 2026
- Establishment of PCOS-specific prolactin reference intervals to optimize laboratory demand management for macroprolactin screeningClinical Biochemistry · 2026