386 False Hook Effect in Prolactin Testing Due to Reagent-Associated Signal Interference on the Siemens Atellica Analyzer
Regina Tirbeni, Wieslaw Furmaga
Key points
From the paper's conclusion- Conclusion All patient samples exhibited systematic increasing prolactin values with dilution, inconsistent with a true hook effect but attributable to reagent-associated background signal.
Abstract
Introduction/Objective A patient with a pituitary mass underwent prolactin testing to evaluate for hook effect on the Siemens Atellica analyzer. Serial dilutions demonstrated progressive increases rather than the expected rise-and-fall pattern of a true hook effect. This study aimed to determine the cause of this atypical behavior and its implications for assay performance and validation. Methods/Case Report Three patient samples were analyzed for prolactin using the Siemens Atellica immunoassay platform. Automated and manual serial dilutions were performed using the manufacturer’s diluent, reagent-grade water, and saline. Reproducibility was assessed across three analyzers. Reagent blanks were tested at room temperature, 37 °C, and 37 °C with solid phase to assess potential background signal and assay interference. Results The initial sample prolactin concentration rose from 1.0 ng/mL undiluted to 2.92 ng/mL with 1:5 automated dilution; manual dilutions (1:5–1:3000) increased to 1443.61 ng/mL without decline. A second sample showed similar increases across diluents (41–968 ng/mL). A third sample reproduced findings across three analyzers. Diluent alone remained low (0.46–0.48 ng/mL), while reagent blanks showed intrinsic signal under all conditions (0.71–1.24 ng/mL). Conclusion All patient samples exhibited systematic increasing prolactin values with dilution, inconsistent with a true hook effect but attributable to reagent-associated background signal. Awareness of this phenomenon is essential during revalidation of new reagent lots and when establishing maximal dilution and hook effect protocols to prevent misinterpretation in suspected hook effect cases.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishment of Sex-Specific Reference Intervals for Total and Post-PEG Prolactin on an Automated Analyzer.The journal of applied laboratory medicine · 2026
- Comparative assessment of cross-reactivity with Big-prolactin and macroprolactin in five commercial assays for prolactin.Clinical chemistry and laboratory medicine · 2026
- Analytically accurate, clinically inaccurate: A diagnostic case of Pseudohyperparathyroidism.Clinica chimica acta; international journal of clinical chemistry · 2026
- Post-Ejaculatory Prolactin Elevation Resolves Within 2 Hours: Implications for Blood Sampling in Male Infertility Evaluation.Andrology · 2026
- Optimization of PEG precipitation for macroprolactinemia screening.Clinical chemistry and laboratory medicine · 2026
- Does Hyperprolactinemia Lead to Lipid Peroxidation or Vice Versa?Journal of obstetrics and gynaecology of India · 2026