Garbage in, (mis)diagnosis out: making laboratory measurement bias visible and actionable across clinical algorithm outputs.
Niels van Noort, Marith van Schrojenstein Lantman, Mathie P G Leers, Remy J H Martens, Volkher Scharnhorst, Marc H M Thelen, Arjen-Kars Boer
SKML, Nijmegen, The Netherlands. · Department of Clinical Chemistry and Hematology, Zuyderland Medical Center, Sittard-Geleen, The Netherlands.
Key points
From the paper's conclusion- By translating bias-propagation analysis into interpretable laboratory reports, this framework bridges the gap between measurand-level analytical performance and clinical algorithm performance. It supports post-deployment monitoring, cross-laboratory comparability and verification, targeted quality improvement, and is applicable to any transparent formula-based clinical algorithm.
Abstract
OBJECTIVES: Clinical algorithms increasingly combine multiple laboratory measurands to support diagnosis, prognosis, and treatment decisions. Yet, measurement bias arising from differences in calibration, reagent lots, or analytical platforms can propagate through such algorithms in non-linear ways, shifting outputs and altering clinical classifications. Existing external quality assessment (EQA) programmes evaluate measurands individually and do not capture the cumulative effect of bias on algorithm behaviour. We present a framework that extends EQA to quantify and communicate the impact of laboratory-specific measurement bias on clinical algorithm outputs. METHODS: Bias estimates derived from EQA results are injected into a representative patient cohort and propagated through the algorithm; when clinical outcomes are available, output shifts are translated into changes in clinically interpretable performance characteristics. Measurand-specific what-if analyses estimate the expected benefit of correcting individual bias. The framework is illustrated using the Fibrosis-4 (FIB-4) index and scalability is demonstrated on the ten-input CoLab COVID-19 severity score. RESULTS: Participant-facing reports were developed that summarise deviation from a zero-bias reference, position each laboratory within the peer distribution, and identify which input measurands offer the greatest potential for performance recovery. Applied to FIB-4, the reports demonstrate how laboratories can locate themselves relative to clinically desired performance thresholds and receive actionable, measurand-specific guidance to prioritise corrective efforts. CONCLUSIONS: By translating bias-propagation analysis into interpretable laboratory reports, this framework bridges the gap between measurand-level analytical performance and clinical algorithm performance. It supports post-deployment monitoring, cross-laboratory comparability and verification, targeted quality improvement, and is applicable to any transparent formula-based clinical algorithm.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Translating analytical performance into clinical misclassification risk: a probabilistic framework for HbA1c interpretation.Clinical chemistry and laboratory medicine · 2026
- PathEQA: Feature-Graph-Guided Random Forests for Multianalyte External Quality AssessmentmedRxiv · 2026
- Combined total error in calculated clinical chemistry tests: component analytical error is magnified and decision threshold gray zones widen.Laboratory medicine · 2026
- Inter-laboratory variability in vancomycin measurement in an external quality assessment scheme 2016–2025Clinical Biochemistry · 2026
- Observed, Curated, Reused: Observation Bias in Clinical AI EvaluationarXiv · 2026
- An improved approach to modelling patient reclassification: HbA 1c as an exampleClinical Chemistry and Laboratory Medicine (CCLM) · 2026