An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of gentamicin in human serum.

Jo Anne Wright, Neeraj Singh, Annika Brauch, Friederike Bauland, Andrea Geistanger, Judith Taibon, Elie Fux

Roche Diagnostics GmbH, Penzberg, Germany. · Evidenze Germany GmbH, Essen, Germany.

Key points

From the paper's conclusion
  • This candidate RMP provides a superior metrological foundation for vancomycin quantification by establishing direct SI-traceability and a fully characterized uncertainty budget. While its implementation necessitates a re-evaluation of clinical decision thresholds, it significantly enhances the global standardization of TDM assays.

Abstract

OBJECTIVES: Therapeutic drug monitoring (TDM) of gentamicin is critical to ensure efficacy and minimize toxicity. We report a candidate reference measurement procedure (RMP) to quantify gentamicin in biological matrices, establishing an unbroken traceability chain to the International System of Units (SI). METHODS: An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method was validated using quantitative nuclear magnetic resonance (qNMR) spectroscopy to replace historical biological activity-based standardization with gravimetric mass. Strict ID-MS principles were applied to the major congeners (C1, C1a, and C2). Isomers C2a and C2b, eluting at distinct retention times, were quantified using C2-D3 as a structural analog internal standard. Pharmacokinetic-based analytical performance specifications set the maximum allowable standard measurement uncertainty (k=1) at ≤6.7 %. RESULTS: The RMP demonstrated high analytical control, with trueness (bias 1.0-2.8 %) and within-run precision (CV 0.2-0.9 %) fulfilling clinical requirements. Expanded measurement uncertainties (k=2) for target value assignment ranged from 2.0 to 2.4 %. Comparison with the current JCTLM-listed RMP revealed a systematic negative bias of -9.2 %. A three-way method correlation confirmed these differences as a direct result of the metrological transition from activity to SI-traceable mass units. CONCLUSIONS: This candidate RMP provides a robust, highly precise SI-traceable foundation for gentamicin quantification. While the resulting downward shift in reported concentrations may necessitate re-evaluating established clinical decision thresholds and TDM protocols, the clinical requirement for such adjustments remains to be fully determined.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of gentamicin in human serum — Science Skim