科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ MethodsX2026-06-09· Chromatography

A robust dilute-and-inject ion chromatography - Tandem mass spectrometry for urinary glyphosate quantification

Mari-Vorgan Louyer, Cécile JABALOT, Arthur David, Fabien Mercier

原始摘要(英文原文)· Original abstract
Glyphosate is one of the most widely used herbicides in the world, whose potential effects on human health are still the subject of debate today. To assess the associated risks, reliable biomonitoring data are required using validated robust and sensitive analytical methods. In this context, an ion chromatography coupled with tandem mass spectrometry (IC-MS/MS) method was developed to quantify glyphosate in human urine. The method is based on direct injection of diluted urine samples. A labeled internal standard ( 13 C 2 15 N 1 glyphosate) is added prior to injection to correct for losses or matrix effects during the analysis. The method limit of quantification (LOQ) is 0.1 ng/mL in urine. The working range was validated between 0.1 and 20 ng/mL. Method trueness was assessed at three concentration levels (LOQ, 2 and 20 ng/mL), and a full validation of the method was performed to characterize the analytical performance in terms of precision and measurement uncertainty. • Direct injection ion chromatography coupled with tandem mass spectrometry • Validation of the method according to international standards
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

A robust dilute-and-inject ion chromatography - Tandem mass spectrometry for urinary glyphosate quantification — 科研速览 Science Skim