Robust ghost marker rejection in marker-based optical tracking using contour-derived monocular depth priors.
Enzan Chen, Rougang Zhou, Ange Wang, Chou Jay Tsai Chien, Junjie Zhu, Zhenchao Ruan
School of Mechanical Engineering, Hangzhou Dianzi University, Hangzhou, China. · Department of Geriatrics, the First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.
Key points
From the paper's conclusion- We developed a tracking method that combines contour-based marker detection and constrained ellipse fitting with contour-derived monocular depth priors, uncertainty-normalized stereo matching, and adaptive region-of-interest (ROI) processing.
Abstract
Marker-based stereo optical tracking can be disrupted by reflective interference and ambiguous correspondences when marker projections lie near the same epipolar line. We developed a tracking method that combines contour-based marker detection and constrained ellipse fitting with contour-derived monocular depth priors, uncertainty-normalized stereo matching, and adaptive region-of-interest (ROI) processing. In the reprojection-consistency evaluation, the proposed localization method achieved mean errors of 0.0532 and 0.0552 pixels in the left and right cameras, respectively, compared with 0.0827 and 0.0843 pixels for grayscale-centroid localization. In a separate inter-marker distance experiment involving 150 stereo observations, its mean absolute error was 0.0458 mm, compared with 0.0515 mm for OpenCV fitEllipse and 0.0551 mm for the grayscale-centroid method. In 50 stereo frames with deliberately induced ghost-marker ambiguity, the proposed matching method met the predefined frame-level correctness criterion in all 50 frames, compared with 20 frames for epipolar-only matching and 37 frames for MSRC. All expected markers were detected without false positives in the 100 images acquired for the paired illumination experiment. With 1-15 markers, adaptive ROI processing maintained a mean algorithmic processing time below 20 ms per stereo frame pair; at 15 markers, the mean was 16.52 ms and the 95th percentile was 21.20 ms. These results demonstrate improved measurement and matching performance under the tested laboratory conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- LARK: A Low-Cost, Accurate, Occlusion-Resilient, Kalman Filter-Assisted Tracking System for Image-Guided SurgeryarXiv · 2026
- Feature image extraction and gaze estimation based on head-mounted eye trackerEngineering Research Express · 2026
- Quantitative Evaluation of OpenCV Visual Tracking Algorithms for Drawn Markers Under Dynamic Facial Deformation.Sensors (Basel, Switzerland) · 2026
- Accuracy of Stereo Endoscope Calibrations for Forward and Oblique OpticsCurrent Directions in Biomedical Engineering · 2026
- AMP: An Automated Marker Based Motion Capture Processing Pipeline for Kinematic Trajectory Label Scrubbing and Gaps FillingbioRxiv · 2026
- Subaperture stitching measurement for optical components based on the trust region reflective algorithm.Optics express · 2026