Current Directions in Biomedical Engineering

Accuracy of Stereo Endoscope Calibrations for Forward and Oblique Optics

Linus L. Kienle, Sophie Beckmann, Niklas Freund, Anna Hilsmann, Peter Eisert, Eric L. Wisotzky

Key points

From the paper's conclusion
  • We present and evaluate calibration methods for a stereo endoscope with interchangeable 0∘ and 30∘ optics.

Abstract

Accurate geometric calibration is a prerequisite for metric 3D reconstruction from stereo endoscopes, but practical deployment is complicated by non-standard optics (e.g., angled views), strong lens distortion, and clinically realistic image artifacts such as specularities and low texture. We present and evaluate calibration methods for a stereo endoscope with interchangeable 0∘ and 30∘ optics. We compare (i) a target-based calibration using an AprilGrid-style planar pattern and (ii) a reconstruction-based calibration approach that optimizes geometric consistency from image correspondences. To quantify calibration quality, we measure known distances on two reference objects. Our results demonstrate how calibration repeatability and optical-center stability directly impact the reliability of intraoperative distance measurements. Furthermore, this study provides practical recommendations on strategies to improve the reliability of intraoperative metric measurements.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers