Three-hospital study of endoscope protein contamination and microbiome.
Anant Agrawal, Sangwon Lee, Alex Chediak, Jahan Azizi, Mohamed E Labib, Kaumudi Kulkarni, Miranda Gavette, Jennifer White,
United States Food and Drug Administration, Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, Division of Biomedical Physics, 10903 New Hampshire Avenue, Silver Spring, Maryland 20993, USA. · United States Food and Drug Administration, Center for Devices and Radiological Health, Office of Product Evaluation and Quality, Office of Health Technology 7, Division of Microbiology Devices, 10903 New Hampshire Avenue, Silver Spring, Maryland 20993, USA.
Key points
From the paper's conclusion- Both methods are suitable for clinical use. For TFF, 68 samples across three hospitals had an average pre-cleaning protein mass of 541±342µg and post-cleaning mass of 59±57 µg. For QC-ER, 63 samples across three hospitals had an average pre-cleaning mass of 66±52µg and post-cleaning mass of 28±33µg. None of 60 samples tested positive for microbes after high-level disinfection.
Abstract
BACKGROUND: Over 15 million colonoscopies and 7 million upper-GI endoscopies are performed annually in the United States, requiring reprocessing after each use to reduce infection risk. Methods for endoscope quality control are limited and require rapid turnover and high sensitivity/specificity. METHODS: Here we explore two novel approaches to measure protein soil on endoscopes: turbulent fluid flow (TFF) extraction and extractionless assay for quick check of endoscope reprocessing (QC-ER). Microbial strains in extracted samples are characterized with Next-gen sequencing and PCR. RESULTS: Both methods are suitable for clinical use. For TFF, 68 samples across three hospitals had an average pre-cleaning protein mass of 541±342µg and post-cleaning mass of 59±57 µg. For QC-ER, 63 samples across three hospitals had an average pre-cleaning mass of 66±52µg and post-cleaning mass of 28±33µg. None of 60 samples tested positive for microbes after high-level disinfection. DISCUSSION: In method validation, both high level disinfection exposure and clinical sample matrices interfered with protein quantification by the bicinchoninic acid (BCA) assay. TFF and QC-ER detected soil that remained after simple rinsing. Differences in clinical results for pre-cleaning were likely due to post-procedure protocols. All levels were below consensus thresholds. PRINCIPAL CONCLUSIONS: The amount of protein on endoscopes in healthcare facilities was quantified by these methods and fell within the expected range, with variations by type and age of endoscope.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study.PloS one · 2026
- A PBMC-based assay detects residual pro-inflammatory response in endoscopes after high-level disinfection undetected by ATP bioluminescence: a pilot study.The Journal of hospital infection · 2026
- Residual protein assessment in medical device processing: from analytical sensitivity to measurement system reliability.The Journal of hospital infection · 2026
- Benchmarking the Intratumoral Microbiome in Pancreatic Ductal Adenocarcinoma: A Longitudinal Assessment of Contamination Sources and Decontamination StrategiesbioRxiv · 2026
- Antibiotic resistance of bacteria isolated from microbiological control of gastrointestinal endoscopes.The Journal of hospital infection · 2026
- Single‐target contamination modeling in pathogen metagenomic sequencing via simple linear regressionVIEW · 2026