Standardizing wrist x-rays: The human-centered design and evaluation of a wrist x-ray positioning tool.
Laura Vancer, Maxwell Campbell, Glenn Schurmans, John Henry, Jamie Kueneman, Joy MacDermid, Gregory Garvin, Emily Lalone
School of Biomedical Engineering, Western University, London, Ontario, Canada. · Roth McFarlane Hand and Upper Limb Centre, St. Joseph's Health Care, London, Ontario, Canada.
Key points
From the paper's conclusion- Factors that affect x-ray positioning and shape device adoption were determined from the preliminary end-user interviews. Through post-intervention questionnaires, the Tru-Vu tool and training modules were found to be easy to use/understand and perceived as effective by radiographers. There was no significant difference observed in overall radiograph quality and visualization of the scapholunate (SL) interval during cadaveric feasibility testing.
Abstract
INTRODUCTION/BACKGROUND: Small amounts of malrotation can significantly decrease the diagnostic utility of a wrist radiograph. Poorly positioned wrist radiographs can result in repeat scans, additional imaging, and missed injuries. Tolerance for mistakes must be low. Radiographers, especially those with limited training, experience, or resources, may struggle to meet an acceptable standard, underscoring the need for an assistive positioning device that makes proper positioning more accessible. METHODS: A human-centered design (HCD) process was used during the design of the Tru-Vu, a low-cost novel positioning aid, to assist radiographers with properly positioning of the wrist for x-rays. Initially, observations and preliminary interviews were conducted to understand the challenges faced by various end-users. The identified themes were used to guide the iterative design of the tool and training modules. To evaluate feasibility, the tool and training modules underwent testing using a cadaveric specimen. RESULTS: Factors that affect x-ray positioning and shape device adoption were determined from the preliminary end-user interviews. Through post-intervention questionnaires, the Tru-Vu tool and training modules were found to be easy to use/understand and perceived as effective by radiographers. There was no significant difference observed in overall radiograph quality and visualization of the scapholunate (SL) interval during cadaveric feasibility testing. DISCUSSION: This study provides insight into the HCD process used to create a novel radiographic positioning aid. The Tru-Vu tool was positively perceived by end-users; however, it did not demonstrate a significant effect on radiograph quality in cadaveric feasibility testing. Further research with live patients is required to demonstrate the clinical effectiveness of the Tru-Vu.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a chest radiography positioning education system using digitally reconstructed radiograph (DRR)-based visual feedback.Radiological physics and technology · 2026
- X-ray image quality before and after radiographer commenting: A quantitative visual grading characteristics analysis.Radiography (London, England : 1995) · 2026
- Ergonomics in Facial Ultrasound: Potential Influence on Image Quality, Reproducibility, and Ultrasound-Guided Aesthetic Procedures.Diagnostics (Basel, Switzerland) · 2026
- Automated Detection of Positioning Errors in Bitewing Radiographs Using Deep LearningClinical and Experimental Dental Research · 2026
- Integrating Extended Reality into Interventional Radiology: A Single Center Feasibility Study of a Virtual Ruler for PICC Line Placement.Journal of vascular and interventional radiology : JVIR · 2026
- New Mixed-Reality Navigation System for Cup Positioning in Trapeziometacarpal Arthroplasty: An In Vitro Validation Study.The journal of hand surgery Asian-Pacific volume · 2026