Using a Stifneck Select CollarTM for hands-free semiautomatic blood flow measurements: A user study.
Reinhard Fuchs, Nathalie Sumrah, Maximilian N Möbius-Winkler, Georg Stachel, Michael Schultz, Ulrich Laufs, Thomas Neumuth, Michael Unger,
Innovation Center Computer Assisted Surgery (ICCAS), Medical Faculty, Leipzig University, Leipzig, Germany. · Klinik und Poliklinik für Kardiologie, Universitätsklinikum Leipzig, Leipzig, Germany.
Key points
From the paper's conclusion- We present a sensor attachment solution usable for semiautomatic ultrasonic (US) Doppler measurements.
Abstract
The percentage of long-term survival in out‑of‑hospital cardiac arrest cases is remarkably low. One approach would be to increase the effectiveness of cardiopulmonary resuscitation (CPR), which is currently not measurable in a quantifiable way. The most significant challenge in providing a mobile solution for CPR evaluation is a mobile, hazard‑free sensor attachment with high usability. We present a sensor attachment solution usable for semiautomatic ultrasonic (US) Doppler measurements. Components are attached to a Stifneck Select Collar™ (Laerdal). An inflatable cushion (TR-Band™, Terumo) allows adjustable contact pressure. A clinical study was conducted in which the system was evaluated based on comfort, pain, sensor support, the viability of Doppler signals, and the absence of skin irritations. The system was utilized in a prospective study involving 102 healthy probands. On a scale between 1 (Low) and 10 (Intense), ratings were 1.19 (SD 0.46), 6.52 (SD 1.78), and 9.95 (SD 0.32) for pain, comfort, and support, respectively. The average duration of application was 31.19 minutes (SD 16.75 minutes). Audible Doppler signals were achieved in 92.2% of the probands, and Doppler curve evaluation was usable in 73.5%. No skin irritations were observed. A hands‑free sensor attachment for a US probe was developed that caused no significant complaints by healthy study volunteers. Medical users assessed its attachment as robust. With its adjustable positioning and easy attachment, the Stifneck modification can form a basis for a mobile US Doppler device, capable of evaluating carotid artery flow during CPR.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating hands-free carotid Doppler ultrasound for real-time return of spontaneous circulation detection in in-hospital cardiac arrest: a multicenter randomized controlled studyResuscitation · 2026
- Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation.bioRxiv : the preprint server for biology · 2026
- A randomised controlled trial of defibrillation with manual pressure augmentation during out-of-hospital cardiac arrestResuscitation · 2026
- Wearable devices for detection of cardiac arrest: a prospective proof-of-concept studyEuropean Heart Journal · 2026
- Femoral Artery Compression Ultrasound as an Adjunct to Clinical Return of Spontaneous Circulation Assessment During Cardiopulmonary ResuscitationJournal of Ultrasound in Medicine · 2026
- Wristband-based automated cardiac arrest detection and alerting: Algorithm performance in relation to the left ventricular functionEuropean Heart Journal · 2026