Quantitative analysis of free polyethylene glycols in PEGylated peptides using reverse-phase high-performance liquid chromatography and charged aerosol detection.
Yue Sun, Xinyue Hu, Yue Huang, Yuqin Yang, Xiaoming Zhang, Yi Li, Ping Lyu, Jing Li
National Institutes for Food and Drug Control, State Key Laboratory of Drug Regulatory Science, NMPA Key Laboratory for Quality Research and Evaluation of Chemical Drugs, Beijing, China. · Beijing Institute of Petrochemical Technology, Beijing, China.
Key points
From the paper's conclusion- Herein, systematically optimized reverse-phase high-performance liquid chromatography (RP-HPLC) coupled with charged aerosol detection (CAD) was developed and validated for the quantification of free PEG in three representative PEGylated peptide therapeutics: PEG-loxenatide, pegmolesatide, and visepegenatide.
Abstract
PEGylation is a well-established strategy used to enhance the pharmacokinetic and pharmacodynamic properties of peptide drugs. However, free polyethylene glycol (PEG) remains a critical process-related impurity. Conventional detection methods for nonchromophoric compounds such as PEG, which include the use of evaporative light scattering (ELS) and refractive index (RI) detectors, are limited by their low sensitivity and poor compatibility with the gradient elution. Herein, systematically optimized reverse-phase high-performance liquid chromatography (RP-HPLC) coupled with charged aerosol detection (CAD) was developed and validated for the quantification of free PEG in three representative PEGylated peptide therapeutics: PEG-loxenatide, pegmolesatide, and visepegenatide. The proposed method implements a co-optimization strategy for CAD parameters, with the power function value (PFV) applied during CAD signal acquisition and the power law(PL) setting adjusted during data processing.This dual optimization effectively linearizes the inherently nonlinear CAD response across a broad concentration range, enabling a linear relationship between PEG concentration and detector response. Method validation demonstrated satisfactory specificity, precision, linearity, accuracy, robustness, and solution stability in accordance with the applicable guidelines and pharmacopoeial requirements. Practical applicability was confirmed by analyzing multiple drug batches and accelerated stability samples.A Wilcoxon signed-rank test showed no statistically significant difference between the results obtained under the default and co-optimized settings (P > 0.05), indicating that the two settings produced comparable results under the conditions evaluated. The proposed HPLC-CAD method provides a practical analytical approach in the analysis of PEGylated peptide pharmaceuticals by achieving a linearized CAD response through PFV and PL co-optimization, providing an effective approach for free PEG quantification in routine quality control.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Site-specific PEGylation of proteins: Insights into structural and functional changesActa Pharmaceutica Sinica B · 2025
- Pegylation modulates ApoE association and hepatic accumulation of DC-Chol-based nanoparticles through steric exclusion.International journal of pharmaceutics · 2026
- Serum Peptides Penetrate Dense poly(ethylene glycol) Coatings on Nanoparticles and Modulate Serum Protein Adsorption.Acta biomaterialia · 2026
- The Ubiquitous Use of Polyethylene Glycol in Pharmaceutical Design and Development: Technological Aspects and Future PerspectivesNanomaterials · 2025
- MS3 acquisition and μ-SPE cleanup enable sensitive quantification of PPG1000 in zebrafish tissues: systematic comparison with MRM and in-source CID strategies.The Analyst · 2026
- Effective Reduction of Anti-PEG Antibody Recognition via PEG Copolymers with Ethyl Glycidyl Ether and Isopropyl Glycidyl Ether.Biomacromolecules · 2026