Adverse event reporting and regulatory analysis of pediatric cardiac assist devices.
Songyining Ding, Susmitha Wunnava, Gaite Liu, Hongyi Yu, Sheng Hu, Yaohua Li, Xinyan Zhang, Peiming Zhang,
School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China. · Harvard-MIT Center for Regulatory Science, Harvard Medical School, Boston, Massachusetts, United States of America.
Key points
From the paper's conclusion- This study provides a descriptive evaluation of reporting trend disparities between pediatric and adult cohorts. Given the inherent constraints of passive surveillance data, these exploratory findings should be interpreted as a reference to guide future prospective evaluations and standard optimizations, rather than immediate, broad regulatory mandates.
Abstract
BACKGROUND: Due to differences in anatomy, coagulation and neurodevelopment, the complication rate of pediatric cardiac assist devices during use is much higher than that of adults, which cannot be directly extrapolated from adult data. Moreover, there are deficiencies in the device supervision process. METHOD: In this study, 280 adverse events of patients under 18 years old and 1,231 adverse events of adult patients were extracted from the FDA MAUDE database from 2015 to 2025. The pediatric complication spectrum was systematically analyzed, and adult data was compared with pediatric data. Combined with literature and typical case analysis, the limitations of traditional databases were jointly revealed. Finally, in accordance with the existing testing standards, the risks exposed after the product is launched on the market are summarized and the regulatory loopholes are revealed. RESULT: MAUDE pediatric reports are scarce and variables are missing, making it impossible to directly calculate the incidence rate. The research calls for the establishment of a multi-source real-time monitoring and pediatric exclusive registration system to preliminarily assess and generate hypotheses regarding the risk of mechanical circulatory support in children. CONCLUSION: This study provides a descriptive evaluation of reporting trend disparities between pediatric and adult cohorts. Given the inherent constraints of passive surveillance data, these exploratory findings should be interpreted as a reference to guide future prospective evaluations and standard optimizations, rather than immediate, broad regulatory mandates.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Adverse Event Profiles of Leadless Pacemaker Systems Reported to the FDA MAUDE Database (2021-2026).Journal of cardiovascular electrophysiology · 2026
- Current Global Trends and Challenges of Pediatric Ventricular Assist Device Programs: An ACTION Collaborative Survey.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Adverse Event Landscape of Pulsed Field Ablation Catheters: A Computational Review of FDA MAUDE Post-Market Surveillance Data.Reviews in cardiovascular medicine · 2026
- Post-market individual-report phenotypes after pulsed-field ablation for atrial fibrillation in a food and drug administration database.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Catheter-Directed Interventional Devices for Pulmonary Embolism: A Maude Database Analysis of Adverse Events.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Utilization of Microaxial Mechanical Circulatory Support Among the Pediatric Population: A Systematic Review.Artificial organs · 2026