Rapid, low-cost colorimetric capillary-driven immunoassay for screening of bacterial STIs: Chlamydia, syphilis, and gonorrhea.
Joowon Park, Elijah J O Barstis, Jonah Rodasta, Brian J Geiss, David S Dandy, Charles S Henry
Department of Chemistry, Colorado State University, CO, 80523, USA. · Department of Microbiology, Immunology and Pathology, Colorado State University, CO, 80523, USA; School of Biomedical and Chemical Engineering, Colorado State University, CO, 80523, USA.
Key points
From the paper's conclusion- Here, we present a quadplex capillary-driven immunoassay (CaDI) for instrument-free, simultaneous detection of three bacterial antigens through capillary action, featuring four channels that enable target-specific assay optimization.
Abstract
Bacterial sexually transmitted infections (STIs) caused by Chlamydia trachomatis, Treponema pallidum, and Neisseria gonorrhoeae persist as a significant global health concern with more than 215 million new cases reported in 2020. Low screening rates among sexually active individuals due to stigma, cost, and limited access to clinics result in delayed treatment and transmission of the bacterial STIs. Current gold standard methods such as nucleic acid amplification tests and serological assays require centralized laboratories, trained personnel, and multiple visits to clinics. Here, we present a quadplex capillary-driven immunoassay (CaDI) for instrument-free, simultaneous detection of three bacterial antigens through capillary action, featuring four channels that enable target-specific assay optimization. Nitrocellulose strips are functionalized with capture antibodies to form test lines, along which the horseradish peroxidase-mediated oxidation of TMB results in a blue-purple colorimetric signal when the target antigen is present. This catalytic signal amplification resulted in limits of detection comparable to ELISAs in single-pathogen CaDIs: 2.6, 7.4, and 16.3 ng/mL for Chlamydia trachomatis, Treponema pallidum, and Neisseria gonorrhoeae antigens, respectively. Quadplex CaDI experiments demonstrated excellent specificity and selectivity, with target signals unaffected by the presence of 10 μg/mL non-target proteins and robust protein A-based positive control lines. The quadplex CaDI is cost-effective, does not require active pumps, and is simple to operate, being suitable for point-of-care and resource-limited screening. While clinical validation against PCR remains as future work to establish clinical sensitivity and specificity, the quadplex CaDI suggests an alternative screening method to increase testing frequency, enable prompt treatment, and reduce bacterial STI transmission.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-078 STI Scout®: Chlamydia and Gonorrhea Molecular ID Testing for Home and Resource Constrained SettingsClinical Chemistry · 2026
- A-365 A Rapid Multiplex LAMP Assay for Detection of Chlamydia trachomatis, Neisseria gonorrhoeae, Trichomonas vaginalis, and Fluoroquinolone Resistance in Neisseria gonorrhoeae with Translatability Across Benchtop and Point-of-Care PlatformsClinical Chemistry · 2026
- Colorimetric-Catalytic-SERS Trimodal Immunochromatographic Assay for Multi-Scenario Monitoring of Gastrointestinal Infections.ACS applied materials & interfaces · 2026
- A-104 Compatibility of Copan UBM Transport Medium with Molecular Detection of STI Targets across Different qPCR AssaysClinical Chemistry · 2026
- Paper-based colourimetric biosensor for simultaneous detection of UTI-associated nitrite and β-lactamase-mediated resistance using an iodometric method.Enzyme and microbial technology · 2026
- A-080 Molecular diagnosis and quantitative analysis of Neisseria gonorrhoeae using qPCRClinical Chemistry · 2026