Orthogonal inline microscopy: continuous determination of microparticle size, shape, and concentration via contrast-gated flow imaging.
Raphael Hauer, Christian Hill, Ulrich Hohenester
Institute of Physics, University of Graz, Universitätsplatz 5, 8010 Graz, Austria. · BRAVE Analytics GmbH, Stiftingtalstraße 14, 8010 Graz, Austria.
Key points
From the paper's conclusion- We assess these capabilities using National Institute of Standards and Technology (NIST)-traceable size and concentration standards in the 2 µm to 41 µm range and wastewater samples collected before and after filtration.
Abstract
Flow imaging microscopy (FIM) has emerged as an essential analytical technique for characterizing suspended microparticles in fields ranging from environmental monitoring to pharmaceutical quality control. Orthogonal inline microscopy (OIM), which emerged from the fluidic infrastructure of optofluidic force induction (OF2i), was recently employed upstream of single-particle inductively coupled plasma mass spectrometry to quantify size-dependent transport efficiencies, as reported in [Anal. Chim. Acta1422, 346044 (2026)]. Here, we present OIM as a stand-alone FIM technique and provide a comprehensive description of its optical setup and contrast-gated data evaluation. OIM is distinguished by its sample transport from established FIM instrumentation: the fluid flows through the measurement cell along the optical axis and orthogonally to the object plane of the imaging optics. The camera captures the entire capillary cross-section, and the complete supplied sample volume traverses the object plane. This geometry enables the continuous extraction of absolute particle size distributions, absolute number concentrations, and morphological shape descriptors without volumetric extrapolation. We assess these capabilities using National Institute of Standards and Technology (NIST)-traceable size and concentration standards in the 2 µm to 41 µm range and wastewater samples collected before and after filtration.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative assessment of the size and concentration measurement in a novel flow cytometer.The Review of scientific instruments · 2026
- Dual-scale and dual-channel oblique plane microscopy for multi-resolution imaging of large tissue specimensBiomedical Optics Express · 2026
- Method Development, Interlaboratory Study, and Quality Control Criteria for Flow Imaging Microscopy Analysis of Phytoplankton in FreshwaterAWWA Water Science · 2026
- Experimental studies on microscopic, multiphase transport phenomena: Novel applications of light sheet fluorescence microscopyJournal of Visualization · 2026
- Integrating Flow Imaging Analysis and Single-Particle ICP-TOFMS for Comprehensive Micro- and Nanoplastic Characterization.Analytical chemistry · 2026
- Monitoring Liquid Concentration Changes in Nanofluidic Channels Using Cross-Grating Wavefront Microscopy.ACS nano · 2026