Method Development, Interlaboratory Study, and Quality Control Criteria for Flow Imaging Microscopy Analysis of Phytoplankton in Freshwater
Hunter Adams, Polly Barrowman, Katie Kohoutek, Christiane Hoppe‐Jones, Susheera Pochiraju, Margaret Spoo‐Chupka, Patricia Klonicki, Eric Johnson,
Key points
From the paper's conclusion- This study presents the development of Standard Method 10210 for FIM: method optimization, QC criteria, and a multi‐laboratory validation study (10 laboratories and 13 analysts).
Abstract
Flow imaging microscopy (FIM) is used for rapid characterization of phytoplankton in water; however, standardized quality control (QC) and interlaboratory validation are lacking. This study presents the development of Standard Method 10210 for FIM: method optimization, QC criteria, and a multi‐laboratory validation study (10 laboratories and 13 analysts). Standardized protocols were established using microbead reference materials and applied to source water samples. Interlaboratory testing demonstrated strong reproducibility, with relative standard deviations (RSDs) for non‐fluorescent microbeads analyzed in AutoImage Mode < 10% (30–70 μm) and < 20% (10 μm). Fluorescent microbead RSDs were < 20% (AutoImage Mode) and < 25% (Trigger Mode). Surface water sample RSDs were ≤ 30% in AutoImage Mode across triplicates, days, and laboratories. The study established QC capable of identifying operational issues and demonstrated consistent classification of phytoplankton across laboratories/holding times. This work provides the first validated QC framework for FIM‐based monitoring, enabling reliable, rapid, and standardized assessment of phytoplankton populations for source management and early bloom detection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of particle properties and biovolume estimates from FlowCam and microscopy for kettle hole phytoplankton communitiesLimnology and Oceanography: Methods · 2026
- Orthogonal inline microscopy: continuous determination of microparticle size, shape, and concentration via contrast-gated flow imaging.Optics express · 2026
- The value of flow cytometry for the rapid preparation of standardized suspensions of viable spores from recurrent airborne molds.Journal of microbiological methods · 2026
- Quantitative assessment of the size and concentration measurement in a novel flow cytometer.The Review of scientific instruments · 2026
- OptoSieve: a white-light-sheet flow cytometer for simultaneous color and multispectral fluorescence phenotyping of phytoplankton.Optics express · 2026
- Three-dimensional optical imaging for label-free sensing and classification of microorganisms and microparticles in fluidic environments: a review [Invited].Optics express · 2026