Ultra-thin, forward-viewing OCT endoscope using a rotating angle-cleaved fiber.
Radu-Florin Stancu, Michael Hughes, Manuel J Marques, Carlo Seneci, Ross Henry, Sébastien Ourselin, Lyndon da Cruz, Christos Bergeles,
University of Kent, Applied Optics Group, School of Engineering, Mathematics and Physics, Canterbury, CT2 7NH, UK. · King's College London, Robotics and Vision in Medicine Lab, School of Biomedical Engineering & Imaging Sciences, 1 Lambeth Palace Rd, South Bank, London, SE1 7EU, UK.
Key points
From the paper's conclusion- Here, we report a forward-viewing OCT fiber endoscope that can fit inside a protective tube with an inner diameter of 0.16 mm, enabling an endoscope with an outer diameter of 0.31 mm.
Abstract
While ultra-thin optical coherence tomography (OCT) probes using rotating fibers have been demonstrated in side-viewing configurations, achieving fully miniaturized forward-viewing OCT endoscopy remains challenging because of the need to incorporate a lateral scanning mechanism. Here, we report a forward-viewing OCT fiber endoscope that can fit inside a protective tube with an inner diameter of 0.16 mm, enabling an endoscope with an outer diameter of 0.31 mm. The rotating fiber assembly incorporates a gradient-index fiber lens and an angle-cleaved length of no-core fiber at the distal tip. This causes the beam to exit at an angle to the endoscope axis, such that rotating the fiber using an external rotary coupler generates circular scanning of the beam over the tissue. Lateral and axial resolutions of 25 µm and 11 µm, respectively, are achieved at focus, with a usable imaging depth range of over 2 mm and an image circle circumference of 950 µm at 1 mm from the fiber tip. An exemplar application in vitreoretinal surgery is demonstrated by ex vivo porcine retinal imaging and real-time monitoring of a surgical tool interacting with the tissue.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fully Fiber-Integrated 3D-Printed Probes for Plug-and-Play Endoscopic Optical Coherence Tomography.ArXiv · 2026
- Theranostic Endoscope for Three‐Dimensional OCT‐Guided Ultrafast Laser Ablation of Atherosclerotic LesionsLaser & Photonics Reviews · 2026
- Miniaturized piezoelectric tube actuated catheter (μPTAC) enabling distal actuation for intravascular optical coherence tomography.Ultrasonics · 2026
- Neurointerventional optical coherence tomography using a piezo-driven microprobe.Biomedical optics express · 2026
- Real-Time Intracochlear Distance Sensing and Navigation Using a Dual Fiber Optical Coherence Tomography System.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Ronchi-calibrated bidirectional fast volume-repeat OCT for high-speed 4D imaging.Optics letters · 2026