Non-contact optical coherence elastography with motion-artifact correction and curvature compensation for quantitative assessment of in vivo keratoconus biomechanics
Yirui Zhu, Likai Huang, Liyang Wang, Ningning Luo, Tomas E. Gomez Alvarez Arenas, Jiulin Shi, Xingdao He
Key points
From the paper's conclusion- Here, we introduce an air-coupled ultrasound transducer-based OCE (AcUT-OCE) system that couples motion-artifact correction with curvature-aware compensation, enabling robust in vivo elastographic mapping of the diseased cornea.
Abstract
Keratoconus (KC) is a progressive ectatic disorder in which stromal thinning and apical protrusion culminate in marked biomechanical decompensation of the cornea. Quantifying these changes in vivo is essential for early detection, monitoring of disease progression, and assessment of therapeutic response. However, conventional optical coherence elastography (OCE) is confounded in vivo by physiological-motion artifacts and by geometric bias inherent to the planar-scan model, which fails to honor the cornea's native curvature. Here, we introduce an air-coupled ultrasound transducer-based OCE (AcUT-OCE) system that couples motion-artifact correction with curvature-aware compensation, enabling robust in vivo elastographic mapping of the diseased cornea. To operationalize both corrections, we constructed a preprocessing pipeline that suppresses physiological-motion phase noise and remaps the lateral coordinate according to corneal curvature. In healthy rabbit corneas, the model-derived effective Young's modulus increased from 7.88 ± 2.19 MPa with conventional planar processing to 9.58 ± 2.63 MPa after combined motion-artifact correction and curvature remapping. This difference demonstrates that the processing strategy materially influences the reconstructed wave speed and effective-modulus estimates. Application of the same workflow to collagenase-induced KC and CXL-treated corneas yielded effective-modulus estimates of 1.93 ± 0.64 MPa for KC, 3.56 ± 0.94 MPa after 15 J/cm 2 CXL, and 8.81 ± 2.01 MPa after 30 J/cm 2 CXL. These results demonstrate distinct biomechanical patterns among healthy, KC, and CXL-treated rabbit corneas. Because the study did not include an independent mechanical reference, the corrected values are interpreted as processing-dependent, model-derived estimates rather than validated absolute material properties.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Advances in the application of optical coherence elastography in ophthalmologyJournal of Innovative Optical Health Sciences · 2026
- Spatiotemporal evolution of corneal biomechanical properties after cross-linking based on optical coherence elastography.Frontiers in bioengineering and biotechnology · 2026
- Integration of Mechanical Testing, In Vivo Optical Coherence Elastography and Personalized Finite Element Modeling to Predict Geometrical Outcomes of Corneal Cross-LinkingAnnals of Biomedical Engineering · 2026
- Biomechanical stabilization of the keratoconic cornea: a review of current therapeutic strategies.Frontiers in medical technology · 2026
- Volumetric imaging of Young's modulus and nonlinear elasticity on the micro-scale.Acta biomaterialia · 2026
- Signal Processing of Corneal Biomechanics in Early Keratoconus Detection: A Critical ReviewSignals · 2026