Spectrometer matching for balanced detection in spectral-domain optical coherence tomography by autocorrelation minimization.

Santosh Balakrishnan, Steven G Adie

Nancy E. and Peter C. Meinig School of Biomedical Engineering, Cornell University, Ithaca, New York 14850, USA.

Key points

From the paper's conclusion
  • Precise matching of spectrometers is essential to realizing the full benefits of balanced detection in spectral-domain optical coherence tomography (SD-OCT).

Abstract

Precise matching of spectrometers is essential to realizing the full benefits of balanced detection in spectral-domain optical coherence tomography (SD-OCT). A retrospective approach to match spectrometers, based on the observation that the autocorrelation signal should be zero in a perfectly balanced system, is presented. This approach requires only a single calibration B-scan, does not depend on the source noise or image properties, and can be used either to guide alignment or to obtain balanced results from previously acquired data. The performance of this method is demonstrated on a fiber-based SD-OCT system with a pair of commercial spectrometers and for a super-luminescent diode or supercontinuum laser source.

Read the paper