From the paper's conclusion- Case Report:A 92-year-old man presented with acute mechanical pain, recurrent clicking, and subjective instability 16 years after primary PS TKA following a low-energy flexion-rotation movement.Plain radiographs showed no evidence of periprosthetic fracture, component loosening, or implant failure.Dynamic fluoroscopy during flexion, which reproduced the patient's symptoms, demonstrated abnormal an
Abstract
BACKGROUND Tibial post fracture is an uncommon but likely underrecognized complication of posterior-stabilized (PS) total knee arthroplasty (TKA) that often remains undiagnosed because conventional radiographs may be unrevealing. We report a late tibial post fracture in a previously well-functioning PS TKA, highlighting the role of dynamic fluoroscopy in raising preoperative suspicion and the successful use of isolated polyethylene insert exchange in a carefully selected patient. CASE REPORT A 92-year-old man presented with acute mechanical pain, recurrent clicking, and subjective instability 16 years after primary PS TKA following a low-energy flexion-rotation movement. Plain radiographs showed no evidence of periprosthetic fracture, component loosening, or implant failure. Dynamic fluoroscopy during flexion, which reproduced the patient's symptoms, demonstrated abnormal anterior translation of the femoral component relative to the tibial insert during knee flexion, raising suspicion of cam-post mechanism failure. Revision surgery confirmed a fracture of the polyethylene tibial post with chronic wear at its base; the metallic components remained well fixed, without gross malposition. Isolated polyethylene insert exchange with a thicker insert restored intraoperative stability. At 4.5 months postoperatively, the patient demonstrated complete resolution of mechanical symptoms, with full extension, 100° to 110° of flexion, and preserved varus-valgus and anteroposterior stability. CONCLUSIONS This case illustrates how dynamic fluoroscopy may raise preoperative suspicion of tibial post fracture in selected patients with symptomatic late PS TKA and inconclusive conventional radiographs. In our patient, isolated polyethylene insert exchange was effective after confirmation of preserved component fixation, alignment, and overall stability, supporting its use in appropriately selected cases.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Occult cement plug causing femoral canal obstruction during nailing after TKA: a case report.Frontiers in surgery · 2026
- Are kneeling stress radiography and the posterior sag sign associated with subjective flexion instability after posterior-stabilized total knee arthroplasty?Archives of orthopaedic and trauma surgery · 2026
- Polyethylene Exchange During Retrograde Intramedullary Nailing in the Treatment of Periprosthetic Distal Femur Fractures: A Registry Pilot Analysis.Journal of the American Academy of Orthopaedic Surgeons. Global research & reviews · 2026
- Catastrophic Failure of a Prosthetic Femoral Stem due to Component Malposition: A Case Report.Case reports in orthopedics · 2026
- Primary Total Knee Arthroplasty Using Revision-Type Stems for Distal Femoral Bone Compromise Following Previous Fracture Fixation: A Report of Two Cases and Narrative Review.Cureus · 2026
- Polyethylene Liner Fracture in Touch Trapeziometacarpal Prosthesis: A Case ReportJournal of Hand Surgery Global Online · 2026