Safety and Efficacy of the Extravascular Implantable Cardioverter-Defibrillator (EV-ICD) in Japanese Patients: Final Results From the EV ICD Japan Study.
Mitsuru Wada, Nobuhiro Nishii, Kohei Ishibashi, Christopher Wiggenhorn, Takuto Yamamoto, Kengo Kusano
Department of Cardiovascular Medicine, National Cerebral and Cardiovascular Center Osaka Japan. · Department of Cardiovascular Therapeutics, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Sciences Okayama Japan.
Key points
From the paper's conclusion- The EV-ICD system was safely implanted and effective at terminating both induced and spontaneous ventricular arrhythmias in a small cohort of Japanese patients.
Abstract
BACKGROUND: The extravascular implantable cardioverter-defibrillator (EV-ICD) has proven efficacy and a low complication rate in global populations, but data from Japanese patients is limited. We report the safety and efficacy of the EV-ICD system in a Japanese cohort through 12 months of follow-up. METHODS AND RESULTS: Patients from Japan with a Class I or IIa indication for a single-chamber ICD were enrolled and implanted with the EV-ICD system. Defibrillation success was determined following implant. System programming changes, device electrical performance, freedom from major system- or procedure-related complications through 12 months post-implant, and follow-up status at all visits were assessed. From February to September 2023, 14 patients (12 males, 2 females) were enrolled and successfully implanted across 2 sites (age: 53.6±14.7 years; left ventricular ejection fraction: 52.6±15.6%; secondary prevention: 42.9%). Defibrillation testing was successful in all patients (14/14, 100%). At 2 weeks and 12 months, freedom from major system- and/or procedure-related complications was 92.9% and 85.7%, respectively. One patient received 7 appropriate shocks that were successful in terminating the arrhythmia. There were no inappropriate shocks delivered, and a single inappropriate antitachycardia pacing therapy was delivered through follow-up. CONCLUSIONS: The EV-ICD system was safely implanted and effective at terminating both induced and spontaneous ventricular arrhythmias in a small cohort of Japanese patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-World Experience With the Extravascular Implantable Cardioverter Defibrillator Through 1 Year: The Enlighten Post Approval Registry Study.Circulation · 2026
- The extravascular ICD: Evolving paradigms in sudden death preventionHeart Rhythm · 2025
- Impact of anatomical factors and implantation position on defibrillation threshold in extravascular implantable cardioverter-defibrillatorEuropean Heart Journal · 2026
- Extravascular Implantable Cardioverter-Defibrillators in Children: Real World Evidence.Pediatric cardiology · 2026
- Transient Defibrillation Threshold Elevation Associated With Air in the Generator Pocket After Extravascular ICD ImplantationHeartRhythm Case Reports · 2026
- A Review of Anesthetic Considerations for Subcutaneous and Extravascular Implantable Cardioverter-Defibrillators.Seminars in cardiothoracic and vascular anesthesia · 2026