Inhibitory Effects of Metal Ions in Radiopharmaceuticals on PCR Amplification for Rapid Sterility Tests.
Takahito Ito, Akiko Hachisuka, Katsuhiko Hayashi, Kenji Ohya, Takahiro Ohnishi, Yosuke Demizu, Yoshiro Saito, Takashi Misawa
Division of Organic Chemistry, National Institute of Health Sciences, 3-25-26 Tonomachi, Kawasaki-ku, Kawasaki 210-9501, Japan. · Division of Microbiology, National Institute of Health Sciences, 3-25-26 Tonomachi, Kawasaki-ku, Kawasaki 210-9501, Japan.
Key points
From the paper's conclusion- Addition of the chelating agent 2,2',2″,2‴-(1,4,7,10-Tetraazacyclododecane-1,4,7,10-tetrayl)tetraacetic acid to Lu-containing samples restored amplification activity, suggesting that radiopharmaceuticals in their chelated form are unlikely to inhibit PCR.
Abstract
Most radiopharmaceuticals are administered as injectable preparations and are therefore subject to sterility testing. Sterility testing usually requires a 14-d incubation period, which poses a significant challenge for short-lived radiopharmaceuticals because testing must be performed retrospectively. Advances in science and technology have led to the development of various rapid microbial detection methods, one of which is the nucleic acid amplification technique employing PCR. However, PCR is known to be inhibited by a variety of metal ions, and radiopharmaceuticals frequently contain metal ions as radionuclides or as additives, raising concerns regarding false-negative results. In the present study, PCR was performed in the presence of metal radionuclides and additives (ZnCl2, InCl3, LuCl3, YCl3, SnCl2, GaCl3, CuCl2, TlCl, and ZrOCl2) used in radiopharmaceuticals. Lutetium (Lu) and yttrium (Y) exhibited particularly strong PCR inhibitory ability across all enzyme systems tested, while zirconium (Zr) exhibited strong but markedly system-dependent inhibition, showing negligible inhibition with one of the enzyme systems tested. Addition of the chelating agent 2,2',2″,2‴-(1,4,7,10-Tetraazacyclododecane-1,4,7,10-tetrayl)tetraacetic acid to Lu-containing samples restored amplification activity, suggesting that radiopharmaceuticals in their chelated form are unlikely to inhibit PCR. Furthermore, when we conducted an analysis using a mock sample that closely resembled the actual radiopharmaceutical, we observed PCR inhibition that appeared to be caused by excipients at concentrations higher than those of the radionuclide, suggesting that pretreatment may be necessary depending on the formulation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Determination of 226Ra in 225Ac Active Pharmaceutical Ingredients by ICP-QQQ-MS.ACS omega · 2026
- Gallium-68 and zirconium-89 complexes of DFO-analogs for molecular imaging applications using PET/CT.EJNMMI radiopharmacy and chemistry · 2026
- A Comparative Study of Ultraviolet Radiation and Steam Autoclave as Sterilization Methods of Zinc-Based Materials for In Vitro Assays.ACS biomaterials science & engineering · 2026
- Enabling Irreversible FAP Inhibitors for Routine Clinical Use: Process and Quality Control Validation for [68Ga]Ga-FAPi-mFS and [177Lu]Lu-FAPi-mFS.Pharmaceuticals (Basel, Switzerland) · 2026
- cGMP production of radiometal solutions from solid targets at a university-scale facility.Nuclear medicine and biology · 2026
- Development of a species-specific PCR-lateral flow assay integrated with SYBR Safe for rapid authentication of commercial Polygoni Multiflori Caulis (Ye-Jiao-Teng).Journal of natural medicines · 2026