Multistage preclinical evaluation of a retrievable 3 T MR-resonant inferior vena cava filter and snare loop for MRI-guided interventions.
Denis Gholami Bajestani, C Martin Reich, Jan Birkigt, Christina Mulik, Johannes Kauffold, Haukur Lindberg Sigmarsson, Lisa Wahl, Michael Schmid,
Innovation Center Computer Assisted Surgery, Faculty of Medicine, Leipzig University, Leipzig, 04103, Germany. denis.gholami@medizin.uni-leipzig.de. · Innovation Center Computer Assisted Surgery, Faculty of Medicine, Leipzig University, Leipzig, 04103, Germany.
Key points
From the paper's conclusion- MRI-guided implantation and SNL-assisted retrieval of an MR-resonant IVC filter at 3 T demonstrated feasibility within a multistage preclinical workflow. Phantom and in vivo experiments showed reliable visualization and procedural guidance, supporting further development toward clinically applicable MRI-guided caval interventions.
Abstract
BACKGROUND: Fluoroscopy-guided inferior vena cava (IVC) filter implantation and retrieval involve ionizing radiation and iodinated contrast agents. MRI-guided interventions may provide a radiation-free alternative with direct visualization of filter position and hook accessibility. This study evaluated the feasibility of MRI-guided implantation and snare loop (SNL)-assisted retrieval of a dedicated MR-resonant IVC filter at 3 T. METHODS: MR-resonant devices were assessed in a three-stage protocol. Device visibility was investigated in a saline phantom using balanced steady-state free precession (bSSFP) imaging at 3 T with the IVC filter oriented parallel to the main magnetic field B0 and assessed at different longitudinal-axis rotations and flip angles. MRI-guided IVC filter implantation was performed in a polymer-based vessel phantom. Finally, MRI-guided IVC filter implantation and SNL-assisted retrieval were performed in porcine models (n = 5). RESULTS: MR-resonant signal enhancement enabled reliable visualization of the IVC filter in phantom and in vivo experiments. Eight filters were successfully implanted in five animals, and SNL-assisted retrieval followed by re-implantation was achieved in all retrieval attempts (3/3). Quantitative signal analysis of the saline phantom experiments demonstrated significant effects of flip angle (p < 0.001), rotation angle (p < 0.001), and filter diameter (p = 0.032) on normalized signal intensity, with the strongest signal enhancement observed at a flip angle of 10°. CONCLUSIONS: MRI-guided implantation and SNL-assisted retrieval of an MR-resonant IVC filter at 3 T demonstrated feasibility within a multistage preclinical workflow. Phantom and in vivo experiments showed reliable visualization and procedural guidance, supporting further development toward clinically applicable MRI-guided caval interventions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multistage Preclinical Evaluation of MRI-Guided Stenting for Coarctation of the Aorta using in vitro and Porcine Models.Cardiovascular and interventional radiology · 2026
- Magnetic resonance imaging safety-driven, heat-mitigating design for a magnetic resonance imaging-guided, magnetically actuated robotic catheter.Medical physics · 2026
- Receive-Only Coupled Wireless Radiofrequency Probe for Endocavity MR Imaging Using a pTx System.Magnetic resonance in medicine · 2026
- 3-T MRI in vitro characterisation of a SPION-polymer balloon catheter: a proof-of-concept study.European radiology experimental · 2026
- A Decoupled Parallelogram-Flexure Mechanism for Remote Center of Motion in Magnetic Resonance Imaging-Guided Abdominal InterventionsJournal of Mechanical Design · 2026
- A Guidewire Trajectory Generation Tool for Interventional MRI SimulationsarXiv · 2026