Harmonizing calibrations for multi-center 177Lu dosimetry: insights from the Pan-Canadian PR.21 trial
Sara Kurkowska, Nadine Colpo, Pedro L. Esquinas, Conor Dellar, Wendy Parulekar, Fred Saad, Daniel Juneau, Kim N. Chi,
Key points
From the paper's conclusion- This study developed a practical and reliable calibration approach for a multi-center pan-Canadian 177 Lu-PSMA-617 trial.
Abstract
Introduction The rapid expansion of radiopharmaceutical therapies (RPTs) has increased the need for multi-center clinical trials, yet achieving consistent dosimetry across institutions with different imaging equipment remains challenging. This study developed a practical and reliable calibration approach for a multi-center pan-Canadian 177 Lu-PSMA-617 trial. Specifically, a centralized core laboratory was established to cross-calibrate equipment and support dosimetry calculations across all sites. This approach reduced inter-site variability and minimized the need for radioactive source handling at remote centers. The method is applicable to future RPT clinical trials requiring harmonized dosimetry. Methods The core laboratory radionuclide calibrator was first calibrated against the manufacturer’s (Novartis AG, Basel, Switzerland) reference standard and validated with a high-purity germanium (HPGe) detector calibrated with a NIST-traceable source. Reference radioactive sources were then prepared, documented, and shipped to the participating sites. Sites performed planar and SPECT/CT acquisitions using these sources and adjusted their radionuclide calibrator dial settings to match the core laboratory activity reference. Sites submitted acquisition data to the core laboratory for verification and protocol compliance. The sensitivity and dead-time constant for each SPECT/CT system were calculated centrally. Calibration consistency between manufacturer and the core laboratory was evaluated by comparing reported patient vial activities with measurements from the core laboratory radionuclide calibrator. Results Eleven sites across Canada including a total of 22 radionuclide calibrators and 12 SPECT/CT systems were calibrated using this method. Although the core-laboratory calibrator was independently anchored to a NIST-traceable HPGe measurement (agreement within 0.5%), manufacturer-stated activities of subsequent vials were systematically lower, with discrepancies up to 13% (median 10%) over three years. The mean SPECT/CT sensitivity was 9.23 ± 0.15 cps/MBq for a Siemens camera with a 3/8″ crystal ( n = 5) and 12.54 cps/MBq for a Siemens camera with a 5/8" crystal ( n = 1). The mean sensitivity was 4.91 ± 0.53 cps/MBq for the GE camera with a 3/8" crystal ( n = 6). The mean dead-time constant was 1.08 ± 0.16 µs (Siemens, n = 6) and 2.95 ± 0.75 µs (GE, n = 5). Harmonization reduced inter-site activity variability across all geometries (e.g., point source: SD 10.3% → 1.6%). Conclusion Centralized calibration harmonization using a core laboratory is feasible and effective for multi-center RPT trials. Preparing reference sources centrally and performing centralized dosimetry calculations ensures consistent and comparable measurements across participating sites. Critically, manufacturer-stated activities drifted by up to 13% during the trial; independent verification against a stable, traceable reference is therefore essential for multi-center dosimetry.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An International Study of Factors Affecting Variability of Dosimetry Calculations, Part 6: Lessons Learned, Achievable Variability, and the Road Ahead.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2026
- Dosimetry Data Infrastructure for Individualized Radiopharmaceutical Therapy: International Experience and a Proposed Framework for China.Pharmaceuticals (Basel, Switzerland) · 2026
- Patient Communication and Radiation Safety in [177Lu]Lu-PSMA Radiopharmaceutical Therapy: International Survey of 102 Theranostic Centers.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2026
- Lutetium-177 radiopharmaceuticals in cancer care: progress and challenges.Frontiers in pharmacology · 2026
- Single-Time-Point Time-integrated Activity Estimation for Dosimetry in 177Lu-PSMA Therapy: Evaluation of Analytical and Machine Learning Methods.Clinical nuclear medicine · 2026
- Single‐time‐point renal dosimetry for [ 177 Lu]Lu‐PSMA‐617: Incorporating multi‐cycle population data using nonlinear mixed‐effects modelsMedical Physics · 2026