Performance evaluation of a position-sensitive SiPM-based gamma camera for intraoperative imaging.

Aramis Raiola, Fabio Acerbi, Cyril Alispach, Domenico Della Volpe, Hossein Arabi, Alberto Gola, Habib Zaidi

Départment de Physique Nucléaire et Corpusculaire (DPNC), Université de Genève, quai Ernest-Ansermet 24, 1211, Geneva, Switzerland. aramis.raiola@unige.ch. · Center for Sensor and Devices (SD), Fondazione Bruno Kessler (FBK), Via Sommarive 18, 38123, Trento, Italy.

Computational

Key points

From the paper's conclusion
  • The POSiCS camera combines compactness, ease of use, and reliable performance, making it suitable for RGS. Its interchangeable configurations allow balancing between sensitivity and resolution depending on surgical needs. The results demonstrate millimeter-scale resolution and adequate sensitivity for [Formula: see text], confirming its suitability for intraoperative use. Imaging with [Formula: see text] was also proven possible, although achieved only with the 55 keV and 113 keV photopeaks. These outcomes support the camera's potential for broader clinical deployment and further development toward advanced intraoperative nuclear imaging.

Abstract

BACKGROUND: The POSiCS camera is a handheld and wireless, small field-of-view gamma camera developed for use in radio-guided surgery (RGS), with sentinel lymph node biopsy (SLNB) as its benchmark application. This compact and lightweight detector (about) maps tissues labeled with a radiotracer to highlight target lesions. By enabling intraoperative visualization close to the surgical field, POSiCS aims to reduce invasiveness and operative time, improve localization accuracy, and minimize post-operative complications. METHODS: The performances of two hardware configurations were tested: high-sensitivity and high-resolution, to meet application-specific needs. Tests assessed key parameters, including extrinsic spatial resolution, energy resolution, and sensitivity, as well as shielding effectiveness and response to different gamma energies. Characterization of the camera's sensitivity and energy resolution employed a Flood (flat) source, while the determination of spatial resolution and lateral collimator penetration was performed using a point source. The response to was also investigated, given its growing role in theranostics, by imaging a cavity filled with the radionuclide and evaluating the reconstruction performance of POSiCS across its different gamma-ray emissions.. RESULTS: In direct contact, the spatial resolution was in high-sensitivity mode and in high-resolution mode. System sensitivity at source-detector distance was and, respectively. Both modes showed an energy resolution of about 20% at 140 keV, though the high-resolution collimator exhibited increased scatter due to its larger tungsten content. CONCLUSION: The POSiCS camera combines compactness, ease of use, and reliable performance, making it suitable for RGS. Its interchangeable configurations allow balancing between sensitivity and resolution depending on surgical needs. The results demonstrate millimeter-scale resolution and adequate sensitivity for, confirming its suitability for intraoperative use. Imaging with was also proven possible, although achieved only with the 55 keV and 113 keV photopeaks. These outcomes support the camera's potential for broader clinical deployment and further development toward advanced intraoperative nuclear imaging.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Performance evaluation of a position-sensitive SiPM-based gamma camera for intraoperative imaging — Science Skim