Novel approach to estimate respiratory quotient from the alveolar gas equation: a proof-of-concept study in pigs.
Joakim Hedov, Josefin Elzén Junker, Magnus Hallbäck, Per-Arne Lönnqvist, Jacob Karlsson
Department of Thoracic and Vascular Surgery, Linköping University Hospital, Garnisonsvägen 10, Linköping, 581 85, Sweden. joahe182@student.liu.se. · Thoracic Surgery Center, Karolinska University Hospital, Eugeniavägen 23, Stockholm, 171 64, Sweden.
Key points
From the paper's conclusion- Reversal of the alveolar gas equation enables estimation of RQ with strong correlation to reference measurements but introduces a consistent positive bias. While not directly interchangeable with Douglas bag measurements, this simplified approach may provide a practical, continuous approximation of RQ in settings where conventional metabolic monitoring is not feasible. Additional validation studies are needed to make this method ready for clinical use.
Abstract
BACKGROUND: The respiratory quotient (RQ) reflects metabolic substrate utilization and is traditionally measured using indirect calorimetry techniques, e.g. Douglas bag, which are technically demanding and not always feasible in clinical practice. The alveolar gas equation (AGE) provides a theoretical relationship between oxygen tension, carbon dioxide tension and RQ and may be rearranged to estimate RQ using standard respiratory measurements. This study aimed to validate a novel method for estimating RQ by reversing the alveolar gas equation. METHOD: In this prospective experimental study, twelve anesthetized, mechanically ventilated pigs were studied under steady-state conditions. RQ was determined using two methods: (1) reference measurement via Douglas bag collection using Haldane transformation (RQref), and (2) calculation from the rearranged alveolar gas equation using inspired and expired gas measurements (RQcalc). Agreement between methods was assessed using paired t-tests, Bland-Altman analysis with bootstrap confidence intervals, and linear regression. RESULTS: Mean RQref was 0.877 (95% CI 0.845-0.908), while mean RQcalc was 0.994 (95% CI 0.961-1.026), demonstrating a significant difference (p < 0.001). Bland-Altman analysis showed a bias of 0.117 (95% CI 0.103-0.131) with limits of agreement from 0.066 to 0.167 and a mean percentage error of 5%. A strong correlation was observed between methods (r = 0.895, p < 0.001), with RQcalc explaining 80% of the variance in RQref (R² = 0.801). CONCLUSION: Reversal of the alveolar gas equation enables estimation of RQ with strong correlation to reference measurements but introduces a consistent positive bias. While not directly interchangeable with Douglas bag measurements, this simplified approach may provide a practical, continuous approximation of RQ in settings where conventional metabolic monitoring is not feasible. Additional validation studies are needed to make this method ready for clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement of four metabolic equations for estimating submaximal cycling oxygen uptake with two portable indirect calorimetry systems in untrained young women.BMC sports science, medicine & rehabilitation · 2026
- Comparing RMR-Derived Proxy Indicators and LEAF-Q for Detecting Low Energy Availability Risk in Mixed-Sport Female Athletes.Sports (Basel, Switzerland) · 2026
- A Comparison of Different Models to Compute Cardiac Output by Open-Circuit Inert Gas Wash-In.Medicine and science in sports and exercise · 2026
- Estimation of resting metabolic rate in professional soccer players: A cross-sectional study comparing traditional predictive equations and a preliminary machine learning model against indirect calorimetry.PloS one · 2026
- Race-Neutral Spirometry to Predict Residual Volume in COPD Patients: The Pittsburgh Residual Volume Calculator.Chronic obstructive pulmonary diseases (Miami, Fla.) · 2026
- Whole-Room Indirect Calorimetry in the Study of Obesity, Insulin Resistance and Metabolic Flexibility: A Narrative Review.Nutrients · 2026