Research note on gravity artifact reduction in IMU signals using Madgwick orientation estimation.

Moritz Toppmöller, Urs-Vito Albrecht

Department of Digital Medicine, Medical School OWL, Bielefeld University, Universitätsstr. 25, Bielefeld, 33615, NRW, Germany. · Department of Digital Medicine, Medical School OWL, Bielefeld University, Universitätsstr. 25, Bielefeld, 33615, NRW, Germany. urs-vito.albrecht@uni-bielefeld.de.

Key points

From the paper's conclusion
  • A computational phantom was developed using the PyBullet physics engine to generate paired datasets of gravity-free acceleration and simulated IMU signals (with gravitational artifacts). The Madgwick Attitude and Heading Reference System (AHRS) filter was applied to estimate the sensor orientation, followed by a low-pass-filtered gravity projection to subtract artifacts. The gravity-removed acceleration closely matched the gravity-free reference across all axes. Absolute mean differences were consistently below 0.006 m/s2, and Pearson correlation coefficients exceeded 0.98. Cross-correlation analysis revealed zero temporal lag, while Bland-Altman analysis showed no proportional bias. The highest precision was observed along the gravitational z-axis. These simulation-only results represent an upper bound under idealized conditions. Experimental validation remains necessary before practical deployment.

Abstract

OBJECTIVE: Gravitational artifacts bias Inertial Measurement Unit (IMU) signals, complicating applications such as Seismocardiography (SCG). This study aimed to validate a gravity-artifact reduction algorithm using Madgwick orientation estimation in a controlled, physics-based simulation. RESULTS: A computational phantom was developed using the PyBullet physics engine to generate paired datasets of gravity-free acceleration and simulated IMU signals (with gravitational artifacts). The Madgwick Attitude and Heading Reference System (AHRS) filter was applied to estimate the sensor orientation, followed by a low-pass-filtered gravity projection to subtract artifacts. The gravity-removed acceleration closely matched the gravity-free reference across all axes. Absolute mean differences were consistently below 0.006 m/s2, and Pearson correlation coefficients exceeded 0.98. Cross-correlation analysis revealed zero temporal lag, while Bland-Altman analysis showed no proportional bias. The highest precision was observed along the gravitational z-axis. These simulation-only results represent an upper bound under idealized conditions. Experimental validation remains necessary before practical deployment.

Read the paper