Research note on gravity artifact reduction in IMU signals using Madgwick orientation estimation.
Moritz Toppmöller, Urs-Vito Albrecht
Department of Digital Medicine, Medical School OWL, Bielefeld University, Universitätsstr. 25, Bielefeld, 33615, NRW, Germany. · Department of Digital Medicine, Medical School OWL, Bielefeld University, Universitätsstr. 25, Bielefeld, 33615, NRW, Germany. urs-vito.albrecht@uni-bielefeld.de.
Key points
From the paper's conclusion- A computational phantom was developed using the PyBullet physics engine to generate paired datasets of gravity-free acceleration and simulated IMU signals (with gravitational artifacts). The Madgwick Attitude and Heading Reference System (AHRS) filter was applied to estimate the sensor orientation, followed by a low-pass-filtered gravity projection to subtract artifacts. The gravity-removed acceleration closely matched the gravity-free reference across all axes. Absolute mean differences were consistently below 0.006 m/s2, and Pearson correlation coefficients exceeded 0.98. Cross-correlation analysis revealed zero temporal lag, while Bland-Altman analysis showed no proportional bias. The highest precision was observed along the gravitational z-axis. These simulation-only results represent an upper bound under idealized conditions. Experimental validation remains necessary before practical deployment.
Abstract
OBJECTIVE: Gravitational artifacts bias Inertial Measurement Unit (IMU) signals, complicating applications such as Seismocardiography (SCG). This study aimed to validate a gravity-artifact reduction algorithm using Madgwick orientation estimation in a controlled, physics-based simulation. RESULTS: A computational phantom was developed using the PyBullet physics engine to generate paired datasets of gravity-free acceleration and simulated IMU signals (with gravitational artifacts). The Madgwick Attitude and Heading Reference System (AHRS) filter was applied to estimate the sensor orientation, followed by a low-pass-filtered gravity projection to subtract artifacts. The gravity-removed acceleration closely matched the gravity-free reference across all axes. Absolute mean differences were consistently below 0.006 m/s2, and Pearson correlation coefficients exceeded 0.98. Cross-correlation analysis revealed zero temporal lag, while Bland-Altman analysis showed no proportional bias. The highest precision was observed along the gravitational z-axis. These simulation-only results represent an upper bound under idealized conditions. Experimental validation remains necessary before practical deployment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Calibration-free IMU-based Relative Orientation Estimation under Dynamic Human MotionarXiv · 2026
- Physical calibration and signal periodic updates to synchronize IMUs for 5-day continuous knee flexion tracking.medRxiv : the preprint server for health sciences · 2026
- An IMU-Based Wearable Device Utilizing Error-State Kalman Filtering for Motion CaptureApplied Sciences · 2026
- A Matrix-Based Sensor Fusion Algorithm for Nine-D.o.F. IMU Attitude EstimationApplied Sciences · 2026
- IMU-based upper limb kinematics validation: sensor fusion optimisation, sensor-to-segment alignment, and joint angle accuracy.bioRxiv · 2026
- Long-Term In Situ Stability and Measurement Uncertainty Budgets for Gravity-Based Calibration of Acceleration SensorsSensors · 2026