A jaw-partitioned VMAT strategy for enhancing spatial dose modulation in 3D-GRID and 3D-lattice radiotherapy for bulky tumors: an in silico dosimetric comparison.
Weiwei Wu, Zhiwei Liu, Hui Yin, Jinwei Li, Chuangao Wang
Department of Radiation Safety, China Institute of Atomic Energy, Fangshan, Beijing, 102488, China. · Department of Radiation Oncology, Ganzhou Cancer Hospital, Zhanggong, Ganzhou, Jiangxi, 341000, China.
Key points
From the paper's conclusion- JP-VMAT improved PVDR, reduced GI, and decreased valley-region dose while maintaining peak-dose coverage, thereby enhancing spatial dose modulation in both 3D-GRID and 3D-lattice SFRT. However, these dosimetric gains were achieved at the cost of substantially increased MUs and CM, which may prolong treatment delivery and increase susceptibility to intrafraction motion. The measured JP-VMAT plans met the predefined QA criterion, although comparative delivery accuracy relative to FOJ-VMAT was not evaluated.
Abstract
BACKGROUND: Spatially fractionated radiation therapy (SFRT) produces alternating high-dose peak and low-dose valley regions within tumors and provides a dose-modulation strategy for bulky tumors. This study evaluated the feasibility and dosimetric advantages of a jaw-partitioned volumetric modulated arc therapy (JP-VMAT) technique for improving the peak-to-valley dose ratio (PVDR) in 3D-GRID and 3D-lattice SFRT. METHODS: Twenty-five patients with bulky solid tumors were retrospectively included. Four VMAT-SFRT plans were generated for each patient: GRID_JP, GRID_FOJ, LTV_JP, and LTV_FOJ. All plans were prescribed 18 Gy in a single fraction. JP-VMAT was compared with conventional full-open-jaw VMAT (FOJ-VMAT) in terms of peak-dose coverage, gradient index (GI), PVDR, valley-region dose, organs-at-risk (OARs) dose, monitor units (MUs), and plan complexity metric (CM). Correlations between gross tumor volume (GTV) and the magnitude of dosimetric improvement were analyzed. Patient-specific ArcCHECK quality assurance was performed only for the JP-VMAT plans; FOJ-VMAT plans were not measured. RESULTS: JP-VMAT maintained high-dose coverage of the peak-dose structures in both GRID and LTV plans. Compared with FOJ-VMAT, JP-VMAT significantly reduced GI from 11.18 to 9.48 in GRID plans and from 14.06 to 11.03 in LTV plans (both p < 0.001). JP-VMAT increased PVDR_Dmean and PVDR_D50%; PVDR_D50% increased from 2.68 to 2.86 in GRID plans and from 2.90 to 3.08 in LTV plans (both p < 0.001), accompanied by a reduction in low-to-intermediate valley dose within the GTV. These dosimetric differences became more pronounced as GTV volume increased. OAR effects were anatomy- and beam-path-dependent, indicating selective dose redistribution rather than uniform sparing; tracheal Dmax increased slightly in LTV plans. However, JP-VMAT substantially increased MUs, from 7,975 to 15,844 in GRID plans and from 8,815 to 18,922 in LTV plans, and also increased CM. All 50 JP-VMAT plans achieved gamma passing rates ≥ 95% (3%/2 mm; 10% threshold), FOJ-VMAT plans were not measured, precluding direct comparison of delivery accuracy. CONCLUSIONS: JP-VMAT improved PVDR, reduced GI, and decreased valley-region dose while maintaining peak-dose coverage, thereby enhancing spatial dose modulation in both 3D-GRID and 3D-lattice SFRT. However, these dosimetric gains were achieved at the cost of substantially increased MUs and CM, which may prolong treatment delivery and increase susceptibility to intrafraction motion. The measured JP-VMAT plans met the predefined QA criterion, although comparative delivery accuracy relative to FOJ-VMAT was not evaluated.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Comparative Analysis of Virtual 2D VMAT-Based SFRT and 3D Lattice Radiotherapy: Evaluation of Dosimetric and Therapeutic Outcomes.Journal of biomedical physics & engineering · 2026
- Dosimetric comparison of pipe-joint versus conventional VMAT for lattice radiotherapy in bulky tumors: a retrospective planning study.Radiation oncology (London, England) · 2026
- A novel staggered lattice SFRT approach for the management of nonresponding bulky tumors: A dosimetric study.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Clinical commissioning of a commercial brass GRID collimator for spatially fractionated radiotherapy.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Spatially fractionated radiation therapy (SFRT): a scoping review of dosimetric and biological evidence supporting a valley dose–driven translational frameworkRadiation Oncology · 2026
- RADT-49 Dosimetric Comparison of adjuvant Stereotactic Radiotherapy (SRT) Versus Conventionally Fractionated Volumetric Modulated Arc Therapy (VMAT) in Glioblastoma MultiformeNeuro-Oncology · 2026